Measurement Specialties 9116 user manual Position, W1200, W0C01

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 95
Image 95

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

The Valve Position indexes (ii=12 and ii=0C) each have two states (00/01) that when combined provide four (4) possible states of the C1/C2/C3/C4 internal valves. This “Logical Rotary Switch” with four (4) positions is summarized in the following chart:

 

 

 

 

C1 Energized

 

C2 Energized

 

 

 

 

C2 Not

 

C1 Not

 

 

 

 

(0C=01)

 

(0C=00)

 

C3 Energized

 

 

PURGE

 

LEAK/CHECK

 

C4 Not

 

 

 

 

 

 

position

 

position

 

(12=01)

 

 

 

 

 

 

 

 

 

 

C4 Energized

 

 

CAL/RE-ZERO

 

RUN

 

C3 Not

 

 

 

 

 

 

position

 

position

 

(12=00)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Example:

Send TCP/IP commands to Model 9116 module (via its connected socket) setting the calibration valve to the CAL (or Re-Zero) position:

w1200"

(Set RUN/CAL valve position)

w0C01

(Set CAL position)

Responses (both commands):

“A”

“A”

Note

If the programmer “knows” that the module is already in the RUN/CAL valve mode, the first command above is optional. For more information, see Figures 4.1 through 4.4 in Chapter 4.

Page 86

www.PressureSystems.com

Page 95
Image 95
Measurement Specialties 9116 user manual Position, W1200, W0C01