Measurement Specialties 9116 user manual Module Disassembly, Electronic Circuit Board Replacement

Models: 9116

1 143
Download 143 pages 39.58 Kb
Page 114
Image 114

Pressure Systems, Inc.

Model 9116 User’s Manual

 

 

In the process of performing general maintenance on a module and in printed-circuit board replacement, the following tools may be required:

3/32" and 5/64" Allen-head screwdrivers,

a 3/16" hex wrench,

a needle nose tweezers,

a Phillips-head screwdriver, and

a small adjustable wrench.

5.1.2Module Disassembly

The following procedure should be used to disassemble any model prior to any maintenance.

(1)Place the scanner with its external connectors facing up. With one hand holding the module housing, remove all screws securing the top plate to the module housing. These are located around the outer edge of the top panel of the module housing. The Model 9116 uses twelve (12) Phillips head screws around the top plate outside perimeter.

(2)When all screws have been removed, gently lift the top panel and attached electronics up and out of the housing. All components of the pressure scanner are attached to the top plate and will lift out of the module housing when the top plate is removed. See Figure 5.2. Carefully remove the Viton gasket with the module top plate. In some cases, it may be easier to hold the top plate and turn the module over, lifting the housing off the top panel.

5.1.3Electronic Circuit Board Replacement

The Model 9116 contains three (3) printed circuit boards (PCB); the PC-322 main board, the PC-323 PowerPC Daughter board, and the PC-327 analog board. However, the PC-322 and PC323 boards are normally left attached to each other and are treated as a single assembly.

Page 105

www.PressureSystems.com

Page 114
Image 114
Measurement Specialties 9116 user manual Module Disassembly, Electronic Circuit Board Replacement