Teledyne Model 300EU Connecting The Status Outputs, M300EU, Analog Output Pin Outs, Output #

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 16
Image 16
Analog Output Pin Outs

GETTING STARTED

 

 

 

 

 

 

 

M300EU

 

 

 

 

 

 

 

 

Addendum to M300E/EM Manual P/N 04288

 

 

 

 

 

Table 3-2:

Analog Output Pin Outs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pin

 

Analog Output

 

 

Standard Voltage

 

Current

 

 

 

 

 

 

Output

 

Loop Option

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

A1

 

 

V Out

 

I Out +

 

 

2

 

 

 

 

Ground

 

I Out -

 

 

 

 

 

 

 

 

 

 

3

 

 

A2

 

 

V Out

 

I Out +

 

 

4

 

 

 

 

Ground

 

I Out -

 

 

 

 

 

 

 

 

 

 

 

5

 

 

A3

 

 

NOT USED

 

 

 

 

6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7

 

 

A4

 

 

V Out

 

Not Available

 

 

8

 

 

 

 

Ground

 

Not Available

 

 

 

 

 

 

 

 

 

3.3.3. CONNECTING THE STATUS OUTPUTS

The information found in Section 3.1.2.3 is correct with the following exception:

Table 3-3: Status Output Pin Assignments

Output #

Status

Condition

Definition

 

 

1

SYSTEM OK

On if no faults are present.

2

CONC VALID

On if CO concentration measurement is valid.

 

If the CO concentration measurement is invalid, this bit is OFF.

 

 

 

 

 

3

HIGH RANGE

On if unit is in high range of DUAL or AUTO Range Modes.

4

ZERO CAL

On whenever the instruments ZERO point is being calibrated.

 

 

 

5

SPAN CAL

On whenever the instruments SPAN point is being calibrated.

6

DIAG MODE

On whenever the instrument is in DIAGNOSTIC mode.

7

A-REF

On whenever the instrument in is A-REF mode.

8

 

SPARE

 

 

 

 

 

D

 

EMITTER BUSS

The emitters of the transistors on pins 1-8 are bussed together.

+

 

DC POWER

+ 5 VDC

Digital Ground

The ground level from the analyzer’s internal DC Power Supplies.

8

05831 Rev. 01

Page 16
Image 16
Teledyne Model 300EU Connecting The Status Outputs, M300EU, Analog Output Pin Outs, 3 Status Output Pin Assignments, A-Ref