CALIBRATION PROCEDURES | M300EU |
| Addendum to M300E/EM Manual P/N 04288 |
MANUAL CALIBRATION WITH ZERO/SPAN VALVES
The information found in Section 7.4 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s).
•Set the display to show the STABILITY test function.
•There is no GAS TO CAL step.
MANUAL CALIBRATION CHECKS WITH ZERO/SPAN VALVES
The information found in Section 7.5 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s).
•Set the display to show the STABILITY test function.
•There is no GAS TO CAL step.
CALIBRATION WITH REMOTE CONTACT CLOSURES
The information found in Section 7.6 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s).
NOTE
It is required that the analyzer be forced to perform an
calibration, however this cannot be forced via the analyzer’s digital control inputs.
The
Teledyne Instruments’ APICOM emulator software.
AUTOMATIC ZERO/SPAN CAL/CHECK (AUTOCAL)
The information found in Section 7.7 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s).
•The following table replaces Table
| Table | AUTOCAL Modes |
|
|
|
MODE NAME |
| ACTION |
DISABLED | Disables the Sequence | |
ZERO | Causes the Sequence to perform a zero calibration/check | |
Causes the Sequence to perform a zero and span concentration | ||
| calibration/check |
|
SPAN | Causes the Sequence to perform a span concentration | |
| calibration/check |
|
Causes the analyzer to perform an |
CO CALIBRATION QUALITY
The information found in Section 7.8 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU.
CALIBRATION OF OPTIONAL SENSORS
The following Sections do not apply to the M300EU:
•7.9.1 SECTION O2 SENSOR CALIBRATION PROCEDURE
•7.9.2 CO2 SENSOR CALIBRATION PROCEDURE
42 | 05831 Rev. 01 |