Teledyne Model 300EU Setup Æ More Æ Vars Internal Variables Vars, M300EU, Description, Allowed

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 40
Image 40
6.5.SETUP Æ MORE Æ VARS: INTERNAL VARIABLES (VARS)

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

6.5.SETUP Æ MORE Æ VARS: INTERNAL VARIABLES (VARS)

The information found in Section 6.12 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):

There is no STABIL_GAS variable available under the VARS menu of the M300EU.

The following table supersedes Table 6-14 of the M300E/EM Operators Manual (P/N 04288).

Table 6-3: Test Functions Defined

NO.

VARIABLE

DESCRIPTION

ALLOWED

VALUES

 

 

 

 

 

 

 

 

 

Changes the internal data acquisition system (iDAS) hold-off

 

 

 

time, which is the duration when data are not stored in the

Can be between 0.5

 

 

iDAS because the software considers the data to be

0

DAS_HOLD_OFF

and 20 minutes

questionable. That is the case during warm-up or just after

 

 

 

the instrument returns from one of its calibration modes to

Default=15 min.

 

 

SAMPLE mode. DAS_HOLD_OFF can be disabled entirely

 

 

 

 

 

in each iDAS channel.

 

 

 

 

 

1

 

Allows the user to set the number of significant digits to the

AUTO, 1, 2, 3, 4

CONC_PRECISION

right of the decimal point display of concentration and

Default=AUTO

 

 

stability values.

 

 

 

 

 

 

 

2

 

Dynamic zero automatically adjusts offset and slope of the

 

DYN_ZERO

CO response when performing a zero point calibration

ON/OFF

 

 

during an AutoCal (Chapter 7).

 

 

 

 

 

 

 

Dynamic span automatically adjusts slope and slope of the

 

 

 

CO response when performing a zero point calibration

 

3

DYN_SPAN

during an AutoCal (Chapter 7).

ON/OFF

 

 

 

Note that the DYN_ZERO and DYN_SPAN features are not

 

 

 

allowed for applications requiring EPA equivalency.

 

 

 

 

 

4

 

Adjusts the speed of the analyzer’s clock. Choose the +

-60 to +60 s/day

CLOCK_ADJ

sign if the clock is too slow, choose the - sign if the clock is

 

 

too fast.

 

 

 

 

 

32

05831 Rev. 01

Page 40