M300EU

GETTING STARTED

M300EU

Addendum to M300E/EM Manual P/N 04288

The pneumatic flow of the M300EU is different from the M300E/EM in two ways:

The addition of a high-flow multi-tube Nafiondryer.

An additional gas path that passes the sample gas though a CO scrubber used during the auto- reference cycle.

Heat Sync

GFC Wheel

Housing

SAMPLE CHAMBER

GFC Motor

 

 

 

Figure 3-2: M300EU Internal Gas Flow

3.3. ELECTRICAL CONNECTIONS

The information found in Section 3.1.2 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exceptions.

3.3.1. POWER CONNECTION

The information found in Section 3.1.2.1 of the M300E/EM manual (P/N 04288) is correct for the M300EU.

6

05831 Rev. 01

Page 14
Image 14
Teledyne Model 300EU manual Electrical Connections, Power Connection, 2 M300EU Internal Gas Flow, 05831 Rev