6.2.2. TEST FUNCTIONS

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

6.2.2. TEST FUNCTIONS

The information found in Section 6.2.1 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):

The following table supersedes Table 6-2 of the M300E/EM Operators Manual (P/N 04288).

Table 6-2: Test Functions Defined

Parameter

Display Title

Units

Meaning

 

 

 

The full-scale limit at which the output range of the analyzer’s Analog

RANGE

 

 

Outputs is currently set.

- -

RANGE

PPB, PPM

THIS IS NOT the Physical Range of the instrument. See Section 6.3.2

RANGE1

UGM, MGM

for more information.

 

RANGE2

 

 

If DUAL or AUTO Range modes have been selected, two RANGE functions

 

 

 

will appear, one for each range.

 

 

PPB, PPM

Standard deviation of CO concentration readings. Data points are recorded

Stability

STABIL

every ten seconds using the last 25 data points. This function can be reset

UGM, MGM

to show O2 or CO2 stability in instruments with those sensor options

 

 

 

 

 

installed.

 

 

 

 

CO Measure

MEAS

MV

The demodulated, peak IR detector output during the measure portion of the

CFG Wheel cycle.

 

 

 

CO Reference

REF

MV

The demodulated, peak IR detector output during the reference portion of

the CFG wheel cycle.

 

 

 

 

 

 

The result of CO MEAS divided by CO REF based on readings taken during

Measurement /

MR Ratio

-

the normal sample measurement portion of the A-REFcycle.

Reference Ratio

This ratio is the primary value used to compute CO concentration. The

 

 

 

 

 

value displayed is not linearized.

 

 

 

 

 

 

 

The result of CO MEAS divided by CO REF based on readings

Auto-Reference

AZERO RATIO

-

taken during the zero-reference portion of the A-REF cycle.

This ratio is the used to compute a reference correction factor

Ratio

 

 

for computing the CO concentration. The value displayed is

 

 

 

 

 

 

not linearized.

Sample Pressure

PRES

In-Hg-A

The absolute pressure of the Sample gas as measured by a pressure

sensor located inside the sample chamber.

 

 

 

Sample Flow

SAMPLE FL

cm3/min

Sample mass flow rate as measured by the flow rate sensor in the sample

 

 

 

gas stream,

Sample

SAMP TEMP

°C

The temperature of the gas inside the sample chamber.

Temperature

 

 

 

Bench

BENCH TEMP

°C

Optical bench temperature.

Temperature

 

 

 

Wheel

WHEEL TEMP

°C

GFC wheel temperature.

Temperature

 

 

 

Box Temperature

BOX TEMP

°C

The temperature inside the analyzer chassis.

Oven

OVEN TEMP2

°C

The current temperature of the circulating air inside the M300EU’s

Temperature

 

 

convection oven area.

Photo-detector

 

 

The drive voltage being supplied to the thermoelectric coolers of the IR

Temp. Control

PHT DRIVE

mV

photo-detector by the sync/demod Board.

Voltage

 

 

 

 

 

 

 

 

 

 

 

 

The sensitivity of the instrument as calculated during the last calibration

Slope

SLOPE

-

activity. The SLOPE parameter is used to set the span calibration point of

 

 

 

the analyzer.

 

 

 

The overall offset of the instrument as calculated during the last calibration

Offset

OFFSET

-

activity. The OFFSET parameter is used to set the zero point of the

 

 

 

analyzer response.

Test Channel

TEST

mV

The raw voltage being output on the analyzer’s A4 analog output. Only

Output

appears when the test channel is assigned a function.

 

 

Current Time

TIME

-

The current time. This is used to create a time stamp on iDAS readings, and

by the AUTOCAL feature to trigger calibration events.

 

 

 