This is the default reporting range mode for the analyzer.

M300EU

M300EU OPERATING INSTRUCTIONS

(Addendum to M300EU Manual - P/N 04145)

 

6.3.3.RNGE Æ MODE Æ SNGL: CONFIGURING THE M300EU ANALYZER FOR SINGLE RANGE MODE

The single range mode sets a single maximum range for the both the A1 and A2 analog outputs. If the single range is selected both outputs are slaved together and will represent the same reporting range span (e.g. 0-50 ppm), however their electronic signal levels may be configured for different ranges (e.g. 0-10 VDC vs. 0-.1 VDC.

This Reporting range can be set to any value between 0.1 ppb and 10,000 ppb. To select SINGLE range mode and set the upper limit of the reporting range, press:

NOTE

This is the default reporting range mode for the analyzer.

05514 Rev A1

25

Page 33
Image 33
Teledyne Model 300EU manual M300EU, This is the default reporting range mode for the analyzer, Rev A1