Teledyne Model 300EU Setup Æ More Æ Diag Using The Diagnostics Functions, Diag I/O, Diag Aout

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 41
Image 41
6.6.SETUP Æ MORE Æ DIAG: USING THE DIAGNOSTICS FUNCTIONS

M300EU

M300EU OPERATING INSTRUCTIONS

(Addendum to M300EU Manual - P/N 04145)

 

6.6.SETUP Æ MORE Æ DIAG: USING THE DIAGNOSTICS FUNCTIONS

The information found in Section 6.13 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):

There is no DISPLAY SEQUENCE CONFIGURATION submenu available under the DIAG menu of the M300EU.

The following table supersedes Table 6-15 of the M300E/EM Operators Manual (P/N 04288).

 

Table 6-4:

M300EU Diagnostic (DIAG) Submenus

 

 

 

 

 

 

 

DIAGNOSTIC FUNCTION AND MEANING

Front Panel

 

 

Mode Indicator

 

 

 

 

 

 

SIGNAL I/O: Allows observation of all digital and analog signals in the

DIAG I/O

See

 

instrument. Allows certain digital signals such as valves and heaters

M300E/EM

 

to be toggled ON and OFF.

 

 

Manual

 

 

 

 

 

ANALOG OUTPUT: When entered, the analyzer performs an analog

DIAG AOUT

See

 

output step test. This can be used to calibrate a chart recorder or to

M300E/EM

 

test the analog output accuracy.

 

 

Manual

 

 

 

 

 

ANALOG I/O CONFIGURATION: This submenu allows the user to

 

 

 

configure the analyzer’s four analog output channels, including

 

 

 

choosing what parameter will be output on each channel. Instructions

DIAG AIO

6.7.1

 

that appear here allow adjustment and calibration the voltage signals

 

 

 

 

associated with each output as well as calibration of the analog to

 

 

 

digital converter circuitry on the motherboard.

 

 

 

 

 

 

 

ELECTRIC TEST: The analyzer is performing an electric test. This test

DIAG OPTIC

See

 

simulates IR detector signal in a known manner so that the proper

M300E/EM

 

functioning of the sync/demod board can be verified.

 

Manual

 

 

 

 

 

DARK CALIBRATION: The analyzer is performing a dark calibration

DIAG ELEC

See

 

procedure. This procedure measures and stores the inherent dc

M300E/EM

 

offset of the sync/demod board electronics.

 

Manual

 

 

 

 

 

PRESSURE CALIBRATION: The analyzer records the current output

DIAG PCAL

See

 

of the sample gas pressure sensor. This value is used by the CPU to

M300E/EM

 

compensate the CO concentration.

 

 

Manual

 

 

 

 

 

FLOW CALIBRATION: This function is used to calibrate the gas flow

DIAG FCAL

See

 

output signals of sample gas and ozone supply. These settings are

M300E/EM

 

retained when exiting DIAG.

 

 

Manual

 

 

 

 

 

05514 Rev A1

33

Page 41
Image 41
Teledyne Model 300EU Setup Æ More Æ Diag Using The Diagnostics Functions, Diag I/O, Diag Aout, Diag Aio, Diag Pcal, M300EU