3.4. INITIAL OPERATION OF THE M300EU

M300EU

GETTING STARTED

(Addendum to M300EU Manual - P/N 04145)

 

3.4. INITIAL OPERATION OF THE M300EU

NOTE

The analyzer’s cover must be installed and securely fastened to ensure that the convection oven portion

of the internal layout is capable of properly creating and controlling temperatures of the analyzer’s

optical bench.

3.4.1. STARTUP AND WARM UP OF THE M300EU

The process for starting and warming up the M300EU is identical to that described in Sections 3.2.1 and 3.2.2 of the M300E/EM Operators Manual (P/N 04288) except:

It is best to alloy the M300EU to operate uninterrupted for at least 2 hours to allow the temperature of all areas of the convection oven area to equalize.

3.4.2. WARNING MESSAGES

The information found in Section 3.2.3 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s)

Table 3-4: Possible Warning Messages at Start-Up

MESSAGE

MEANING

 

 

ANALOG CAL WARNING

The instrument's A/D circuitry or one of its analog outputs is not calibrated.

 

 

AZERO WARN 1.001

Auto-reference ratio below the specified limits.

 

 

BENCH TEMP WARNING

Optical bench temperature is outside the specified limits.

 

 

BOX TEMP2 WARNING
The temperature inside the M300EU chassis is outside the specified

 

limits (Replaces BOX TEMP WARNING)

CANNOT DYN SPAN

Remote span calibration failed while the dynamic span feature was set to turned

on

 

 

 

CANNOT DYN ZERO

Remote zero calibration failed while the dynamic zero feature was set to turned on

 

 

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

 

 

DATA INITIALIZED

iDAS data storage was erased.

 

 

FRONT PANEL WARN

Firmware is unable to communicate with the front panel.

 

 

OVEN TEMP
The temperature of the insulated convection oven area of the

WARNING

analyzer is outside of the specified limits.

PHOTO TEMP WARNING

Photometer temperature outside of warning limits specified by

PHOTO_TEMP_SET variable.

 

 

 

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

 

 

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

 

 

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

 

 

SAMPLE PRESS WARN

Sample pressure outside of operational parameters.

 

 

SAMPLE TEMP WARN

The temperature of the sample gas is outside the specified limits.

 

 

SOURCE WARNING

The IR source may be faulty.

 

 

SYSTEM RESET

The computer was rebooted.

 

 

WHEEL TEMP WARNING

The Gas Filter Correlation wheel temperature is outside the specified limits.

 

 

05514 Rev A1

9