Signal Name

M300EU Addendum to M300E/EM Manual APPENDIX A-4: Signal I/O Definitions for 300E/EM Series Analyzers, Rev. K.6

APPENDIX A-4: Signal I/O Definitions for 300E/EM Series Analyzers, Rev. K.6

Table A-7: Additional or Changed Test Signal I/O Definitions for M300EU, Revision K.6

 

Signal Name

 

Bit or Channel

 

Description

 

 

 

 

 

 

 

Number

 

 

 

 

 

 

 

 

 

A status outputs, U24, J1017, pins 18 = bits 07, default I/O address 323 hex

 

 

 

 

 

 

 

 

 

 

ST_AUTO_REF

 

6

0

= in auto-reference mode

 

 

1

= not in auto-reference mode

 

 

 

 

 

ST_CONC_ALARM_2

 

7

0

= conc. limit 2 exceeded

 

 

1

= conc. OK

 

 

 

 

 

A status outputs, U24, J1017, pins 1–8 = bits 0–7, default I/O address 323 hex

 

 

 

 

 

 

 

 

 

ST_AUTO_REF

 

6

0

= in auto-reference mode

 

 

1

= not in auto-reference mode

 

 

 

 

 

 

Relay board digital output (PCF8574), default I2C address 44 hex

 

 

 

 

 

 

 

 

 

OVEN_HTR

 

3

0

= oven heaters on

 

 

1

= off

 

 

 

 

Table A-8: M300E/EM Signal I/O Definitions Not Applicable to the M300EU, Revision K.6

Because the CO2 and O2 sensor options are not available on the M300EU, the following signals listed in Appendix A-4 of the M300E/EM Operators Manual (P/N 04288) are not present in the M300EU software.

 

Signal Name

 

 

Bit or Channel

 

 

Description

 

 

 

 

 

 

 

 

Number

 

 

 

 

 

 

 

 

 

 

 

 

Alarm outputs, U21, J1009, pins 112 = bits 47, default I/O address 325 hex

 

 

 

 

 

 

 

 

 

 

 

 

ST_CONC_ALARM_1

 

5

 

1

= conc. limit 1 exceeded

 

 

 

0 = conc. OK

 

ST_CONC_ALARM_2

 

6

 

1

= conc. limit 2 exceeded

 

 

 

0

= conc. OK

 

 

 

 

 

 

 

A status outputs, U24, J1017, pins 1–8 = bits 0–7, default I/O address 323 hex

 

 

 

 

 

 

 

 

 

 

 

ST_CONC_ALARM_1

 

6

 

0

= conc. limit 1 exceeded

 

 

 

1

= conc. OK

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ST_CONC_ALARM_2

 

7

 

0

= conc. limit 2 exceeded

 

 

 

1

= conc. OK

 

 

 

 

 

 

 

 

Relay board digital output (PCF8574), default I2C address 44 hex

 

 

 

 

 

 

 

 

 

 

 

O2_CELL_HEATER 5

 

3

 

0

= O2 sensor cell heater on

 

 

 

1

= off

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rear board primary MUX analog inputs

 

 

 

 

 

 

 

 

 

 

 

O2_SENSOR

 

10

 

O2 concentration sensor

 

 

 

 

 

 

 

CO2_SENSOR

 

12

 

CO2 concentration sensor

 

 

 

 

 

 

 

 

 

 

Rear board temperature MUX analog inputs

 

 

 

 

 

 

 

 

 

 

 

TEMP_INPUT_5

 

5

 

Diagnostic temperature input

 

 

 

 

 

 

 

 

 

05924 Rev K.6

A-13

Page 89