4.Choose a Test Channel parameter to be output on the channel.

M300EU

M300EU OPERATING INSTRUCTIONS

(Addendum to M300EU Manual - P/N 04145)

 

To configure the analyzer’s three analog outputs, set the electronic signal type of each channel and calibrate the outputs. This consists of:

1.Selecting an output type (voltage or current, if an optional current output driver has been installed) and the signal level that matches the input requirements of the recording device attached to the channel.

2.Determine if the over-range feature is needed and turn it on or off accordingly.

3.If a Voltage scale is in use, a bipolar recorder offset may be added to the signal if required.

4.Choose a Test Channel parameter to be output on the channel.

5.Calibrating the output channel. This can be done automatically or manually for each channel.

To access the analog I/O configuration sub menu, press:

SAMPLE

RANGE=50.000 PPM

CO= XXXX

<TST

TST>

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

 

CFG

DAS

RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Toggle these

keys to enter the

correct

PASSWORD

 

SETUP X.X

SECONDARY SETUP MENU

 

COMM

VARS DIAG AREF

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

ENTER PASSWORD:818

 

 

 

 

 

 

8

1

8

 

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG

 

 

 

SIGNAL I/O

 

 

 

 

 

 

 

NEXT

 

 

 

 

ENTR

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Continue pressing NEXT until ...

AIO Configuration Submenu

DIAG

 

ANALOG I/O CONFIGURATION

 

PREV

NEXT

 

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

A OUTS CALIBRATED: NO

 

<SET

SET>

CAL

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_1: 5V, OVR, NOCAL

<SET

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_2: 5V, OVR, NOCAL

<SET

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

TEST_OUTPUT: 5V,OVR, NOCAL

<SET

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

AIN CALIBRATED: NO

 

<SET

SET>

CAL

EXIT

 

 

 

 

 

 

 

Adjusts the signal output

for Analog Output A1

Adjusts the signal output

for Analog Output A2

Selects the parameter top be output on the TEST channel and adjusts its signal output