3.GETTING STARTED

M300EU

GETTING STARTED

(Addendum to M300EU Manual - P/N 04145)

 

3.GETTING STARTED

3.1. UNPACKING THE M300EU

Unpack the M300EU as per the directions in Section 3.1 of the M300E/EM manual (P/N 04288), with the exception that there are no shipping screws.

3.2. M300EU ANALYZER LAYOUT

The front panel of the M300EU is identical to that of the M300E/EM (see Figure 3-1 of the M300E/EM Operators Manual (P/N 04288). The Rear Panel is also very similar to that of the M300E/EM, the only difference being that the instrument’s particulate filter is mounted externally on the upper left side of the rear panel rather than internally as on the M300E/EM.

On the other hand, the internal layout of the M300EU is quite different from the M300E/EM. Most of these differences are related to the need to create a thermally insulated, convection-heated oven in which the optical bench temperature is raised and maintained at a high and very stable temperature. Additionally there is a multi- tube, high flow Nafiondryer that removes moisture from the sample gas.

On/Of Switch &

GFC Wheel

 

GFC Wheel Housing

Circuit Breaker

IR Source

Motor

& Heat Sync

 

Front

Oven

Panel

Temperature

 

Sensor

Optional

 

Shutoff Valve

 

Optional

Oven Heater 1

Zero /Sample/Cal

 

Valves

 

Internal

 

Pump

 

+12 VDC

AREA

Power Supply

INSULATED OVEN

±5 VDC

Power Supply

 

Relay PCA

Oven

Fan

 

AC Power

 

Receptacle

 

(under fan)

 

REAR

 

Panel

Motherboard

 

Stabilization Fan

Main

Gas inlets &

Cooling Fan

Outlets

OPTICAL BENCH

SYNC/DEMOD PCA

Gas

Pressure

Sensor

Gas

Flow

Sensor

Nafion

Dryer

Auto-Ref Valve

& CO Scrubber

Assembly

Fan

3.1. UNPACKING THE M300EU Oven Heater 2

Optional

Ethernet

PCA

CPU

3.2. M300EU ANALYZER LAYOUT Motherboard

Particulate Filter

Figure 3-1: M300EU Internal Layout

05514 Rev A1

5

Page 13
Image 13
Teledyne Model 300EU manual Getting Started, UNPACKING THE M300EU, 3.2. M300EU ANALYZER LAYOUT, Optical Bench, Rev A1