Teledyne Model 300EU SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4, M300EU, Test Channel, Zero

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 46
Image 46
6.7.3. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

6.7.3. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4

This section replaces Section 6.13.5.5 of the M300E/EM Operators Manual (P/N 04288)

The Test Functions available to be reported on analog output A4 are:

Table 6-6:

Test Channels Functions Available on the M300EU’s Analog Output

 

 

 

 

TEST CHANNEL

DESCRIPTION

ZERO

FULL SCALE

 

 

 

 

 

NONE

 

TEST CHANNEL IS TURNED OFF

 

 

 

 

 

 

 

 

The raw output of the optical bench’s IR

 

 

CO MEAS

 

detector during the measure phase of the m/r

0 mV

5000 mV*

 

 

cycle

 

 

CO REF

 

The raw output of the optical bench’s IR

 

 

 

detector during the reference phase of the

0 mV

5000 mV*

 

 

m/r cycle

 

 

SAMPLE PRESSURE

The pressure of gas in the optical bench’s

0 "Hg

40 "Hg-In-A

sample chamber

 

 

 

 

SAMPLE FLOW

 

The gas flow rate through the optical bench’s

0 cm3/min

6000 cm3/min

 

 

sample chamber

 

 

SAMPLE TEMP

 

The temperature of gas in the optical bench’s

0 C°

70 C°

 

sample chamber

 

 

 

 

BENCH TEMP

 

The temperature of optical bench’s itself

0 C°

70 C°

 

 

 

 

 

WHEEL TEMP

 

The temperature of GFC wheel

0 C°

70 C°

 

 

 

 

 

 

 

The temperature of the circulating air inside

 

 

OVEN TEMP

 

the convection oven section of the M300EU’s

0 C°

70 C°

 

 

interior.

 

 

PHT DRIVE

 

The drive voltage being supplied to the

0 mV

5000 mV

 

thermoelectric coolers of the IR photo-detector by

 

 

the sync/demod Board.

 

 

TEMP4

 

SPARE

 

 

 

 

 

 

 

Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds TEST to the list of Test Functions viewable via the Front Panel Display.

38

05831 Rev. 01

Page 46
Image 46
Teledyne Model 300EU SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4, M300EU, Test Channel, Description, Zero, Full Scale