Teledyne Model 300EU manual Specifications, Approvals And Warranty, M300EU, Rev A1

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 11
Image 11
2.SPECIFICATIONS, APPROVALS AND WARRANTY

M300EU

SPECIFICATIONS, APPROVALS AND WARRANTY

(Addendum to M300EU Manual - P/N 04145)

 

2.SPECIFICATIONS, APPROVALS AND WARRANTY

2.1. SPECIFICATIONS

 

 

Table 2-1:

M300EU Basic Unit Specifications

 

 

 

Ranges

 

User selectable to any full scale range from 0-100 ppb to 0-100 ppm

Measurement Units

 

ppb, ppm, µg/m3, mg/m3 (user selectable)

Zero Noise

 

≤ 10 ppb RMS1, 6

Span Noise

 

< 0.5% of reading RMS over 5 ppm1, 3

Lower Detectable Limit

 

< 20 ppb1

 

Zero Drift (24 hours)

 

< 20 ppb2

 

Zero Drift (7 days)

 

< 20 ppb2

 

Span Drift (24 hours)

 

< 0.5% of reading 4

Span Drift (7 days)

 

< 1% of reading 2 4

Linearity

 

Better than 1% of Range5

Precision

 

0.5% reading1, 5

Lag Time

 

<10 sec1

 

Rise/Fall Time

 

<30 sec to 95%1

Sample Flow Rate

 

1800 cm3/min. ± 20%

Temperature Range

 

15 - 35°C operating

Humidity Range

 

0-95% RH, Non-Condensing

Voltage Coefficient

 

< 0.05 % of reading per V

Dimensions (HxWxD)

 

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

 

 

 

 

Weight

 

50 lb (22.7 kg)

 

 

 

 

 

AC Power

 

100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A),

 

220 – 240 V 50/60 Hz (2.5A)

 

 

 

Environmental Conditions

 

Installation Category (Over voltage Category) II Pollution Degree 2

Analog Outputs

 

Three (3) Outputs

 

 

 

 

Analog Output Ranges

 

100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated current loop.

 

All Ranges with 5% Under/Over Range

 

 

 

Analog Output Resolution

 

1 part in 4096 of selected full-scale voltage

Status Outputs

 

8 Status outputs from opto-isolators

Control Inputs

 

6 Control Inputs, 2 defined, 4 spare

 

 

 

 

I/O

 

One (1) RS-232; One (1) RS-485/RS-232/Ethernet

 

Baud Rate : 300 - 115200

 

 

 

Certifications

 

CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A

1

As defined by the USEPA

 

 

 

2

At constant temperature and voltage

 

3

Or 10 ppb, whichever is greater

 

4

Or 10 ppb, whichever is greater

 

5

Above 1 ppm range, otherwise 20 ppb for lower ranges

05514 Rev A1