APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6

APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6

M300EU Addendum to M300E/EM Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6

Table A-1: Additional or Changed Setup Variables Relevant to the M300EU, Revision K.6

 

SETUP VARIABLE

 

 

NUMERIC

 

 

 

DEFAULT

 

 

 

VALUE RANGE

 

 

DESCRIPTION

 

 

 

 

 

 

 

 

 

 

UNITS

 

 

 

VALUE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Low

Access Level Setup Variables (818 password)

 

 

 

Same as in the M300E/EM Operators Manual Appendix A (P/N 04906)

 

 

STABIL_GAS

 

 

 

 

CO

 

CO, CO2 , O2

 

CO2 , O2 not applicable to the M300EU

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Medium Access Level Setup Variables (929 password)

 

 

CO_SAMPLE

 

 

Samples

4

 

 

 

1–30

 

Number of samples to take in measure or

 

 

 

 

 

 

 

reference mode.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FILT_SIZE

 

 

Samples

1000

 

 

 

1–1000

 

Moving average filter size.

 

FILT_DELTA

 

 

PPM

0.7

 

 

 

1–1000

Absolute change to trigger adaptive filter.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO_CONST1

 

 

78.8

 

 

 

100–50000

 

CO calculation constant.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO_CONST2

 

 

1.458

 

 

 

0–10

 

CO calculation constant.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BENCH_SET

 

 

ºC

 

48

 

 

 

0–100

 

Optical bench temperature set point and warning

 

 

 

 

 

Warnings:

 

 

 

 

 

 

 

 

 

limits.

 

 

 

 

 

 

 

 

34–53

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHEEL_SET

 

 

ºC

 

62

 

 

 

0–100

 

Wheel temperature set point and warning limits.

 

 

 

 

 

Warnings:

 

 

 

 

 

 

 

 

 

 

57–67

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ON applies auto-reference offset and dilution

 

ZERO_APPLY_IN_CAL

 

 

 

 

ON

 

OFF, ON

 

factor during zero/span calibration;

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OFF disables both

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_DWELL

 

 

Minutes

3

 

 

 

1–60

 

Dwell time after closing or opening zero scrubber

 

 

 

 

 

 

 

valve.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_SAMPLES

 

 

Samples

1000

 

 

 

1–1000

Number of zero samples to average.

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_FILT_SIZE

 

 

Samples

5

 

 

 

1–100

Auto-zero offset moving average filter size.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_LIMIT

 

 

Ratio

1.15

 

 

 

0–5

 

Minimum auto-zero ratio allowed; must be

 

 

 

 

 

 

 

greater than this value to be valid.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_CAL

 

 

Ratio

1.18

 

 

 

0.5–5

Calibrated auto-zero ratio.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO_TARG_ZERO1

 

 

Conc.

0

 

 

 

-100.00–999.99

 

Target CO concentration during zero offset

 

 

 

 

 

 

 

calibration of range 1.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO_TARG_MID1

 

 

Conc.

50

 

 

 

0.01–9999.99

 

Target CO concentration during mid-point

 

 

 

 

 

 

 

calibration of range 1.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL_BOX_TEMP1

 

 

ºC

30

 

 

 

0–100

Calibrated box temperature for range 1.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO_TARG_ZERO2

 

 

Conc.

0

 

 

 

-100.00–999.99

 

Target CO concentration during zero offset

 

 

 

 

 

 

 

calibration of range 2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO_TARG_MID2

 

 

Conc.

50

 

 

 

0.01–9999.99

 

Target CO concentration during mid-point

 

 

 

 

 

 

 

calibration of range 2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL_BOX_TEMP2

 

 

ºC

30

 

 

 

0–100

Calibrated box temperature for range 2.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Same as

 

 

 

 

TEST_CHAN_ID

 

 

 

 

NONE

 

M300E/EM with

 

Diagnostic analog output ID.

 

 

 

 

 

 

following addition:

 

 

 

 

 

 

 

 

 

 

 

 

 

TEMP4

 

 

 

 

SAMP_FLOW_SET

 

 

cc/m

 

1800

 

 

 

0–5000

 

Sample flow warning limits. Set point is not used.

 

 

 

 

 

Warnings:

 

 

 

 

 

 

 

 

 

 

1400–2500

 

 

 

 

 

 

 

 

 

 

 

 

 

46

 

 

 

 

 

 

Internal box temperature #2/oven set point and

 

OVEN_SET

 

 

ºC

 

 

Warnings:

 

0–100

 

 

 

 

 

 

 

 

warning limits.

 

 

 

 

 

 

 

 

41–51

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OVEN_CYCLE

 

 

Seconds

10

 

 

 

0.5–30

 

Internal box temperature #2/oven control cycle

 

 

 

 

 

 

 

period.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table continues

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 