Teledyne Model 300EU Relay Board Status Led’S, M300EU, Voltage displayed should change. If not

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 72
Image 72
11.2.2. RELAY BOARD STATUS LED’S

TROUBLESHOOTING & REPAIR

M300EU

 

Addendum to M300E/EM Manual P/N 04288

11.2.2. RELAY BOARD STATUS LED’S

The information found in Section 11.1.4.3 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception:

The following table replaces Table 11-5 of the M300E/EM Operators Manual.

Table 11-3: Relay Board Status LED Failure Indications

LED

FUNCTION

SIGNAL I/O PARAMETER

DIAGNOSTIC TECHNIQUE

ACTIVATED BY

VIEW RESULT

 

 

 

 

 

 

 

Voltage displayed should change. If not:

D2

WHEEL

 

 

Failed Heater

WHEEL_HEATER

WHEEL_TEMP

Faulty Temperature Sensor

Yellow

HEATER

 

 

Failed AC Relay

 

 

 

 

 

 

 

 

Faulty Connectors/Wiring

 

 

 

 

Voltage displayed should change. If not:

D3

BENCH

 

 

Failed Heater

BENCH_HEATER

BENCH_TEMP

Faulty Temperature Sensor

Yellow

HEATER

 

 

Failed AC Relay

 

 

 

 

 

 

 

 

Faulty Connectors/Wiring

 

 

 

 

Voltage displayed should change. If not:

D4

OVEN

 

 

Failed Oven Heater(s)

OVEN_HEATER

OVEN_TEMP

Failed Oven Fans(s)

Yellow

HEATERS

Faulty Oven Temperature Sensor

 

 

 

 

 

 

Failed AC Relay

 

 

 

 

Faulty Connectors/Wiring

 

 

 

 

Sample/Cal Valve should audibly change states. If

 

SAMPLE/CAL

 

 

not:

D5

 

 

Failed Valve

GAS VALVE

CAL_VALVE

N/A

Failed Relay Drive IC on Relay Board

Green

OPTION

 

 

Failed Relay Board

 

 

 

 

 

 

 

Faulty +12 VDC Supply (PS2)

 

 

 

 

Faulty Connectors/Wiring

 

 

 

 

Zero/Span Valve should audibly change states. If

 

ZERO/SPAN

 

 

not:

D6

SPAN_VALVE

 

Failed Valve

GAS VALVE

N/A

Failed Relay Drive IC on Relay Board

Green

OPTION

 

 

Failed Relay Board

 

 

 

 

 

 

 

Faulty +12 VDC Supply (PS2)

 

 

 

 

Faulty Connectors/Wiring

 

 

 

 

Shutoff Valve should audibly change states. If not:

D7

SHUTOFF

 

 

Failed Valve

SHUTOFF_VALVE

N/A

Failed Relay Drive IC on Relay Board

Green

VALVE OPTION

Failed Relay Board

 

 

 

 

 

 

Faulty +12 VDC Supply (PS2)

 

 

 

 

Faulty Connectors/Wiring

 

 

 

 

Voltage displayed should change. If not:

 

 

 

 

Failed IR Source

D8

 

IR_SOURCE

CO_MEASURE

Faulty +12 VDC Supply (PS2)

IR SOURCE

Failed Relay Board

Green

 

 

 

Failed IR Photo-Detector

 

 

 

 

 

 

 

 

Failed Sync/Demod Board

 

 

 

 

Faulty Connectors/Wiring

64

05514 Rev A1

Page 72
Image 72
Teledyne Model 300EU manual Relay Board Status Led’S, M300EU, 3 Relay Board Status LED Failure Indications, Ovenheater