Teledyne Model 300EU Auto-Reference Mode Aref, Mode Field, M300EU, AUTO-REF RANGE = 50.000 PPM

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 29
Image 29
6.2.1. AUTO-REFERENCE MODE (AREF)

M300EU

M300EU OPERATING INSTRUCTIONS

(Addendum to M300EU Manual - P/N 04145)

 

6.2.1. AUTO-REFERENCE MODE (AREF)

One of the most significant differences between the M300E/EM and the M300EU analyzers is the auto-reference measurement feature. In this mode, the analyzer makes special measurements and calculations that are applied to the CO concentration calculation to dramatically improve interferent rejection as well as compensate t for changes in ambient temperature of the sample gas and age related drift of the optical bench components (see Section 10.1.1 for detailed information about how and when this A-REFfeature occurs).

When in A-REFmode, the analyzer:

Freezes the CO concentration reading displayed on the front panel and output via the analog outputs or COM Ports.

Displays the Message AUTO-REFin the mode filed of the analyzer’s front panel.

Sets the A-REFstatus output (pin-7 on the status output connector) to high.

Mode Field

AUTO-REF RANGE = 50.000 PPM

 

CO =12.430
<TST TST> CAL CALZ CALS

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 6-1: Typical Front Panel Display during A-REF Mode

NOTE:

Initiating a calibration through either the keyboard, the COMM ports or digital control inputs terminates

the A-REF mode.

Also, when the instrument is in calibration mode, the A-REF mode is suppressed until the instrument

exits calibration mode.

See Section 6.4 for information about changing the A-REFcycle time or manually initiating an auto-reference measurement.

05514 Rev A1

21

Page 29
Image 29
Teledyne Model 300EU Auto-Reference Mode Aref, Mode Field, M300EU, AUTO-REF RANGE = 50.000 PPM, CO =12.430, Setup, Rev A1