Teledyne Model 300EU manual Calibration Of The Analog Outputs, M300EU, Rev A1

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 45
Image 45
6.7.2. CALIBRATION OF THE ANALOG OUTPUTS

M300EU

M300EU OPERATING INSTRUCTIONS

(Addendum to M300EU Manual - P/N 04145)

 

6.7.2. CALIBRATION OF THE ANALOG OUTPUTS

This section replaces Section 6.13.6 of the M300E/EM Operators Manual (P/N 04288).

ANALOG OUPUT calibration needs to be carried out on first startup of the analyzer (performed in the factory as part of the configuration process) or whenever re-calibration is required. The analog outputs can be calibrated automatically or adjusted manually.

During automatic calibration, the analyzer tells the output circuitry to generate a zero mV signal and high-scale point signal (usually about 90% of chosen analog signal scale) then measures actual signal of the output. Any error at zero or high-scale is corrected with a slope and offset.

Automatic calibration can be performed via the AOUTS CALIBRATION command, or by using the CAL button located inside each of the output submenus (e.g. CONC_OUT_1, CONC_OUT_2 or TEST_OUTPUT). By default, the analyzer is configured so that calibration of analog outputs can be initiated as a group with the AOUT CALIBRATION command or individually.

ENABLING OR DISABLING THE AUTO-CAL FOR AN INDIVIDUAL ANALOG OUTPUT

The information found in Section 6.13.5.9 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):

There are minor differences in the displays ( e.g. CONC_OUT_2 instead of DATA_OUT_2)

AUTOMATIC ANALOG OUTPUT CALIBRATION

The information found in Section 6.13.6.2 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU.

MANUAL CALIBRATION OF ANALOG OUTPUT CONFIGURED FOR VOLTAGE RANGES

The information found in Section 6.13.6.3 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):

There are minor differences in the displays ( e.g. CONC_OUT_2 instead of DATA_OUT_2) User Notes:

MANUAL CALIBRATION OF ANALOG OUTPUTS CONFIGURED FOR CURRENT LOOP RANGES

The information found in Section 6.13.6.4 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):

There are minor differences in the displays ( e.g. CONC_OUT_2 instead of DATA_OUT_2)

The current output option can only be installed on outputs A1 and A2 of the M300EU

CALIBRATING THE INTERNAL A TO D CONVERTER (AIN CALIBRATION)

The information found in Section 6.13.6.5 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU.

05514 Rev A1

37

Page 45
Image 45
Teledyne Model 300EU manual Calibration Of The Analog Outputs, M300EU, Automatic Analog Output Calibration, Rev A1