To set the ranges press following keystroke sequence:

M300EU OPERATING INSTRUCTIONS

M300EU

Addendum to M300E/EM Manual P/N 04288

6.3.4.RNGE Æ MODE Æ DUAL: CONFIGURING THE M300EU ANALYZER FOR DUAL RANGE MODE

DUAL range mode allows the A1 and A2 outputs to be configured with separate reporting range spans as well as separate electronic signal levels. The analyzer software calls these two ranges LOW and HI.

The LOW range setting corresponds with the analog output labeled A1 on the rear panel of the instrument and is viewable via the test function RANGE 1.

The HIGH range setting corresponds with the A2 output and is viewable via the test function RANGE 2.

While the software labels these two ranges as LOW and HI, when in DUAL mode their upper limits need not conform to that convention. The upper span limit of the LOW/RANGE1 can be a higher number than that of HI/RANGE 2

To set the ranges press following keystroke sequence:

SAMPLE

 

RANGE=50.000 PPM

CO= XXXX

 

<TST

TST>

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

 

 

 

CFG

DAS

RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE MODE MENU

 

 

 

 

MODE SET

UNIT

 

 

 

EXIT

 

 

SETUP X.X

RANGE MODE:SNGL

 

 

 

 

SNGL DUAL AUTO

 

 

 

ENTR EXIT

 

The LOW and HIGH

SETUP X.X

RANGE MODE:SNGL

 

 

 

 

 

 

 

 

 

 

 

ranges have separate

SNGL DUAL AUTO

 

 

 

ENTR EXIT

 

slopes and offsets for

 

 

 

 

 

 

 

 

computing the CO

 

 

 

 

 

 

 

 

concentration.

SETUP X.X RANGE MODE MENU

 

 

 

 

 

 

 

The two ranges must

MODE

SET

UNIT

 

 

 

EXIT

 

be independently

 

 

 

 

 

 

 

 

calibrated.

 

 

 

 

 

 

 

 

 

SETUP X.X LOW RANGE:500.0 Conc

 

 

 

0

0

5

0

0

.0

ENTR EXIT

EXIT discards the

 

 

 

 

 

 

 

 

new setting

 

 

 

 

 

 

 

 

ENTR accepts the

 

 

 

 

 

 

 

 

new setting

 

SETUP X.X HIGH RANGE:500.0 Conc

 

 

 

0

0

5

0

0

.0

ENTR EXIT

 

Toggled these keys to

 

 

 

 

 

 

 

EXIT discards the new

select the upper

 

 

 

 

 

 

 

 

 

 

 

 

 

 

setting

SPAN limit for the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

reporting range

 

 

 

 

 

 

 

ENTR accepts the

 

 

 

 

 

 

 

 

.

 

 

 

 

 

 

 

new setting

 

 

 

 

 

 

 

 

26

05831 Rev. 01

Page 34
Image 34
Teledyne Model 300EU manual M300EU, To set the ranges press following keystroke sequence, 05831 Rev