Teledyne Model 300EU M300EU Addendum to M300E/EM Manual, A-10, Rev K.6, O2FILTPCT, STDO2CELLTEMP

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 86
Image 86
O2_FILT_PCT

 

APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6

M300EU Addendum to M300E/EM Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP VARIABLE

 

 

NUMERIC

 

 

 

DEFAULT

 

 

 

VALUE RANGE

 

DESCRIPTION

 

 

 

 

UNITS

 

 

 

VALUE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_FILT_PCT

%

 

2

 

 

 

 

0.1–100

Relative change in O2 concentration to

 

 

 

 

 

 

 

shorten filter.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_FILT_DELAY

 

Seconds

20

 

 

 

 

0–300

Delay before leaving O2 adaptive filter mode.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_DIL_FACTOR

 

1

 

 

 

 

0.1–1000

Dilution factor for O2. Used only if is dilution

 

 

 

 

 

 

 

enabled with FACTORY_OPT variable.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO2_COMP_ENABLE

 

 

 

OFF

 

 

ON, OFF

ON enables CO2 compensation; OFF disables

 

 

 

 

 

 

 

it.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO2_COMP_CONC

%

 

0

 

 

 

 

0–20

CO2 concentration to compensate for.

 

 

O2_CELL_SET

 

ºC

 

50

 

 

 

 

30–70

O2 sensor cell temperature set point and

 

 

 

 

 

Warnings:

 

 

 

 

 

 

 

 

 

warning limits.

 

 

 

 

 

 

 

 

 

45–55

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STD_O2_CELL_TEMP

 

ºK

323

 

 

 

 

1–500

Standard O2 cell temperature for temperature

 

 

 

 

 

 

 

compensation.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_ENABLE

 

 

 

ON, OFF

 

 

OFF, ON

ON enables auto-zero calibration using

 

 

 

 

 

 

 

scrubber; OFF disables it.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ZERO_FREQ

 

Minutes

5

 

 

 

 

0.1–1440

Auto-zero calibration period.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO2_TARG_SPAN_CONC

%

 

12

 

 

 

 

0.1–1000

Target CO2 concentration during span

 

 

 

 

 

 

 

calibration.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO2_SLOPE

 

1

 

 

 

 

0.5–5

CO2 slope.

 

 

CO2_OFFSET

%

 

0

 

 

 

 

-10–10

CO2 offset.

 

 

O2_TARG_SPAN_CONC

%

 

20.95

 

 

 

 

0.1–100

Target O2 concentration during span

 

 

 

 

 

 

 

calibration.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_SLOPE

 

1

 

 

 

 

0.5–2

O2 slope.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_OFFSET

%

 

0

 

 

 

 

-10–10

O2 offset.

 

 

CO2_RANGE

%

 

15

 

 

 

 

0.1–500

CO2 concentration range.

 

 

O2_RANGE

%

 

100

 

 

 

 

0.1–500

O2 concentration range.

 

 

O2_CELL_CYCLE

 

Seconds

10

 

 

 

 

0.5–30

O2 cell temperature control cycle period.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_CELL_PROP

 

1

 

 

 

 

0–10

O2 cell PID temperature control proportional

 

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_CELL_INTEG

 

0.1

 

 

 

 

0–10

O2 cell PID temperature control integral

 

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2_CELL_DERIV

 

 

 

0 (disabled)

 

 

0–10

O2 cell PID temperature control derivative

 

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A-10

05924 Rev K.6

Page 86
Image 86
Teledyne Model 300EU manual M300EU Addendum to M300E/EM Manual, A-10, Rev K.6, O2FILTPCT, STDO2CELLTEMP