6.2. OPERATING MODES

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

6.2. OPERATING MODES

The information found in Section 6.1 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s).

The following table supersedes Table 6-2 of the M300E/EM Operators Manual (P/N 04288).

 

Table 6-1:

M300EU Operating Modes

 

 

 

MODE

 

MEANING

 

The analyzer is currently recording values for CO MEAS and CO ref, while the

A-REF

sample gas stream is being routed through the CO scrubber of the auto-

 

reference gas path.

 

DIAG

One of the analyzer’s diagnostic modes is being utilized.

M-P CAL

This is the basic, multi-point calibration mode of the instrument and is activated by pressing

the CAL key.

 

 

 

SAMPLE

Sampling normally, flashing indicates adaptive filter is on.

SAMPLE A

Indicates that unit is in SAMPLE Mode and AUTOCAL feature is activated.

SETUP1

SETUP mode is being used to configure the analyzer (CO sampling will continue during this

 

process as well as data collection and output).

SPAN CAL A2

Unit is performing span cal procedure initiated automatically by the analyzer’s AUTOCAL

 

feature.

 

SPAN CAL M2

Unit is performing span cal procedure initiated manually by the user.

SPAN CAL R2

Unit is performing span cal procedure initiated remotely via the RS-232, RS-4485 or digital i/o

 

control inputs.

 

ZERO CAL A2

Unit is performing zero cal procedure initiated automatically by the analyzer’s AUTOCAL

 

feature.

 

ZERO CAL M2

Unit is performing zero cal procedure initiated manually by the user.

ZERO CAL R2

Unit is performing zero cal procedure initiated remotely via the RS-232, RS-4485 or digital I/O

 

control inputs.

 

1The revision of the Teledyne Instruments software installed in this analyzer will be displayed following the word SETUP. E.g. “SETUP G.5

2The various CAL modes allow calibration of the analyzer. Because of their importance, these modes are described separately in Chapter 7 of the M300E/EM Operators Manual (P/N 04288).