Warning Messages – Indicated Failures (cont.)

 

M300EU

 

 

TROUBLESHOOTING & REPAIR

 

 

(Addendum to M300EU Manual - P/N 04145)

 

 

 

 

Table 11-1:

Warning Messages – Indicated Failures (cont.)

 

 

 

 

 

 

 

 

WARNING

FAULT CONDITION

POSSIBLE CAUSES

 

 

MESSAGE

 

 

 

 

 

 

 

 

Sample flow rate is

Failed sample pump

 

 

 

Blocked sample inlet/gas line

 

 

SAMPLE FLOW

<1400 cm3/min

 

 

WARN

 

or

Dirty particulate filter

 

 

 

Leak downstream of critical flow orifice

 

 

 

 

3

 

 

 

> 2200 cm /min.

Failed flow sensor/circuitry

 

 

 

 

 

 

 

 

 

 

If sample pressure is < 10 in-hg:

 

 

 

Sample Pressure is

o Blocked particulate filter

 

 

 

o Blocked sample inlet/gas line

 

 

 

<10 in-Hg or > 35 in-Hg

 

 

SAMPLE PRES

o Failed pressure sensor/circuitry

 

 

Normally 29.92 in-Hg at sea level

 

 

WARN

decreasing at 1 in-Hg per 1000 ft of

If sample pressure is > 35 in-hg:

 

 

o Pressurized sample gas. Install vent

 

 

 

altitude (with no flow – pump

 

 

 

disconnected).

o Blocked vent line on pressurized sample/zero/span gas

 

 

 

 

 

supply

 

 

 

 

 

o Bad pressure sensor/circuitry

 

 

 

 

 

Ambient temperature outside of specified range

 

 

 

Sample temperature is < 10oC or >

Failed bench heater

 

 

SAMPLE TEMP

Failed bench temperature sensor

 

 

WARN

 

100oC.

Relay controlling the bench heater

 

 

 

 

 

Failed relay board

 

 

 

 

 

I2C buss

 

 

 

 

 

GFC wheel stopped

 

 

 

Occurs when CO Ref is <1250

Failed sync/demod board

 

 

 

If status LED’s on the sync/demod board ARE flashing the

 

 

 

mVDC or >4950 mVDC.

 

 

SOURCE

cause is most likely a failed:

 

 

 

 

 

 

WARNING

Either of these conditions will result

IR source

 

 

 

Relay board

 

 

 

in an invalid M/R ratio.

 

 

 

I2C buss

 

 

 

 

 

 

 

 

 

 

IR photo-detector

 

 

 

 

 

This message occurs at power on. If you have not cycled the

 

 

 

The computer has rebooted.

power on your instrument:

 

 

SYSTEM RESET

o Failed +5 VDC power,

 

 

 

 

 

 

 

 

 

o Fatal error caused software to restart

 

 

 

 

 

o Loose connector/wiring

 

 

 

 

 

Blocked cooling vents below GFC Assembly. Make sure that

 

 

 

 

 

adequate clear space beneath the analyzer.

 

 

 

 

 

Analyzer’s top cover removed

 

 

WHEEL TEMP

The filter wheel temperature is

Wheel heater

 

 

WARNING

controlled at 68 ± 2 °C

Wheel temperature sensor

 

 

 

 

 

Relay controlling the wheel heater

 

 

 

 

 

Entire relay board

 

 

 

 

 

I2C buss

 

05514 Rev A1

61

Page 69
Image 69