Teledyne Model 300EU M300EU, Default Settings, Sngl, Dual, Auto, ÅÆRange1 Low ÅÆ, Low Range

Models: Model 300EU

1 94
Download 94 pages 28 Kb
Page 32
Image 32
DEFAULT SETTINGS

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

The A1 and A2 channels output a signal that is proportional to the CO concentration of the sample gas. They can be configured:

With independent reporting ranges reporting a “single” output signal (SNGL Mode, see Section 6.3.3) o

Be to operate completely independently (DUAL mode, see Section 6.3.4). In this mode the user can set the measure span and signal scale of each output in a variety of combinations (but not the units of measure).

EXAMPLE:

A1 OUTPUT: Output Signal = 0-5 VDC representing 0-1000 ppb concentration values

A2 OUTPUT: Output Signal = 0 – 10 VDC representing 0-500 ppb concentration values.

Or to automatically switch between the two ranges dynamically as the concentration value fluctuates (AUTO modes, see Section 6.3.5).

Both the A1 and A2 outputs can be:

Configured full scale outputs of: 0 - 0.1 VDC; 0 - 1VDC; 0 - 5VDC or; 0 - 10VDC.

Equipped with optional 0-20 mADC current loop drivers and configured for any current output within that range (e.g. 0-20, 2-20, 4-20, etc.).

The user may also add a signal offset independently to each output (see Section 6.7.1) to match the electronic input requirements of the recorder or data logger to which the output is connected.

DEFAULT SETTINGS

The default setting for these the reporting ranges of the analog output channels A1 and A2 are:

SNGL mode

0 to 500.0 ppb

0 to 5 VDC

Reporting range span may be viewed via the front panel by viewing the RANGE test function. If the DUAL or AUTO modes are selected, the RANGE test function d will be replaced by two separate functions, RANGE1 & RANGE2. Reporting range status is also available as output via the external digital I/O status bits.

NOTE

Upper span limit setting for the individual range modes are shared. Resetting the span limit in one

mode also resets the span limit for the corresponding range in the other modes as follows:

SNGL

DUAL

AUTO

Range

ÅÆRange1 (Low)ÅÆ

Low Range

 

Range2 (Hi) ÅÆ

High Range

24

05831 Rev. 01

Page 32