CO2_FILT_PCT

M300EU Addendum to M300E/EM Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP VARIABLE

 

 

NUMERIC

 

 

 

DEFAULT

 

 

 

VALUE RANGE

 

 

DESCRIPTION

 

 

 

 

UNITS

 

 

 

VALUE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OVEN_PROP

 

1/ºC

0.5

 

 

 

0–100

 

Internal box temperature #2/oven PID

 

 

 

 

 

 

proportional coefficient. Proportional band is the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

reciprocal of this setting.

 

OVEN_INTEG

 

0.02

 

 

 

0–100

 

Internal box temperature #2/oven PID integral

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OVEN_DERIV

 

0

 

 

 

0–100

 

Internal box temperature #2/oven PID derivative

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

42

 

 

 

 

 

 

Internal box temperature #2/oven set point and

 

OVEN_SET

 

ºC

 

 

 

 

 

 

0–100

 

 

 

 

 

Warnings:

 

 

 

 

 

 

 

 

warning limits.

 

 

 

 

 

 

 

 

37–47

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BENCH_CYCLE

 

Seconds

15

 

 

 

0.5–30

 

Optical bench temperature control cycle period.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BENCH_PROP

 

 

 

 

 

 

 

 

 

 

 

 

Optical bench temperature PID proportional

 

 

1/ºC

1.5

 

 

 

0–100

 

coefficient. Proportional band is the reciprocal of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

this setting.

 

BENCH_INTEG

 

10.5

 

 

 

0–100

 

Optical bench temperature PID integral

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BENCH_DERIV

 

0

 

 

 

0–100

 

Optical bench temperature PID derivative

 

 

 

 

 

 

coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHEEL_CYCLE

 

Seconds

8

 

 

 

0.5–30

Wheel temperature control cycle period.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHEEL_PROP

 

1/ºC

0.3

 

 

 

0–100

 

Wheel temperature PID proportional coefficient.

 

 

 

 

 

 

Proportional band is the reciprocal of this setting.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHEEL_INTEG

 

0.06

 

 

 

0–100

 

Wheel temperature PID integral coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

WHEEL_DERIV

 

0

 

 

 

0–100

 

Wheel temperature PID derivative coefficient.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table A-2: M300E/EM Setup Variables Not Applicable to the M300EU, Revision K.6

Because the CO2 and O2 sensor options are not available on the M300EU, the following variables listed in Appendix A-2 of the M300E/EM Operators Manual (P/N 04288) are not present in the M300EU software.

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

Medium Access Level Setup Variables (929 password)

CO2_DWELL

Seconds

0.1

0.1–30

Dwell time before taking each sample.

 

 

 

 

 

CO2_FILT_ADAPT

ON

ON, OFF

ON enables CO2 adaptive filter; OFF disables

it.

 

 

 

 

CO2_FILT_SIZE

Samples

48

1–300

CO2 moving average filter size.

 

 

 

 

 

CO2_FILT_ASIZE

Samples

12

1–300

CO2 moving average filter size in adaptive

mode.

 

 

 

 

CO2_FILT_DELTA

%

2

0.01–10

Absolute CO2 conc. change to trigger adaptive

filter.

 

 

 

 

CO2_FILT_PCT

%

10

0.1–100

Percent CO2 conc. change to trigger adaptive

filter.

 

 

 

 

CO2_FILT_DELAY

Seconds

90

0–300

Delay before leaving CO2 adaptive filter mode.

CO2_DIL_FACTOR

1

0.1–1000

Dilution factor for CO2. Used only if is dilution

enabled with FACTORY_OPT variable.

 

 

 

 

O2_DWELL

Seconds

1

0.1–30

Dwell time before taking each sample.

 

 

 

 

 

O2_FILT_ADAPT

ON

ON, OFF

ON enables O2 adaptive filter; OFF disables it.

O2_FILT_SIZE

Samples

60

1–500

O2 moving average filter size in normal mode.

 

 

 

 

 

O2_FILT_ASIZE

Samples

10

1–500

O2 moving average filter size in adaptive

mode.

 

 

 

 

O2_FILT_DELTA

%

2

0.1–100

Absolute change in O2 concentration to

shorten filter.

 

 

 

 

Table continues

 

 

 

 