6.3.6.SETUP Æ RNGE Æ UNIT: SETTING THE REPORTING RANGE UNIT TYPE

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

6.3.6.SETUP Æ RNGE Æ UNIT: SETTING THE REPORTING RANGE UNIT TYPE

The M300EU can display concentrations in ppb, ppm, ug/m3, mg/m3 units. Changing units affects all of the COM port values, and all of the display values for all reporting ranges. To change the units of measure press:

SAMPLE

RANGE=50.000 PPM

CO= XXXX

<TST TST>

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Toggled these keys to

select the units of measure for the reporting ranges

 

SETUP X.X PRIMARY SETUP MENU

 

 

 

 

CFG DAS

RNGE PASS

 

CLK MORE

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE MODE MENU

 

 

 

 

MODE SET

UNIT

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

CONC UNITS:PPB

 

 

 

 

PPB PPM

UGM MGM

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EXIT discards the new

setting

ENTR accepts the new setting Pressure (STP).

NOTE

Concentrations displayed in mg/m3 and ug/m3 use 0°C , 760 mmHg for Standard Temperature and

Pressure (STP).

Consult your local regulations for the STP used by your agency.

NOTE

Once the units of measurement have been changed, the unit MUST be recalibrated, as the “expected

span values” previously in effect will no longer be valid.

Simply entering new expected span values without running the entire calibration routine is not

sufficient.

The following equations give approximate conversions between volume/volume units and

weight/volume units:

CO ppb x 1.25 = CO ug/m3 CO ppm x 1.25= CO mg/m3

28

05831 Rev. 01

Page 36