3.5.3. BASIC CO CALIBRATION PROCEDURE

GETTING STARTED

M300EU

 

Addendum to M300E/EM Manual P/N 04288

3.5.3. BASIC CO CALIBRATION PROCEDURE

The initial calibration should be carried out using the same reporting range set up as used during the analyzer’s factory calibration. This will allow you to compare your calibration results to the factory calibration as listed on the Final Test and Validation Data Sheet.

Because the M300EU’s analog output setup differs from that of the M300E/EM and because the M300EU cannot have be modified to include the optional CO2 or O2 sensor packages, there are some minor differences in the initial Calibration procedure as described in Section 3.3 of the M300E/EM Operators Manual (P/N 04288).

These are:

STEP 1 - SET LIMITS:

Reporting Range Limit should be set to 50.000 ppm.

STEP 2 – DILUTION RATIO:

Ignore this step

STEP 3 – SET CO SPAN GAS CONCENTRATION

The CO span gas concentration should be 40.000 ppm

STEP 4 – ZERO/SPAN CALIBRATION

Set the display to show the STABILITY test function.

There is no GAS TO CAL step because the M300EU cannot be modified to include the O2 and CO2 sensor packages.

Ignore Sections 3.3.2 and 3.3.3 of the M300E/EM Operators Manual (P/N 04288) since they relate to sensor options not available on the M300EU.

USER NOTES: