Visioneer 400 Calibrating from the Windows Control Panel, Double-click on Scanners and Cameras

Models: 400

1 118
Download 118 pages 7.32 Kb
Page 104
Image 104

MAINTENANCE 102

5.Insert the Calibration page in the scanner, in the direction of the arrow(s) on the page.

The calibration page can be inserted in either direction, as long as one of the arrows is pointing in towards the scanner.

6.When calibration is complete the scanner will eject the page out of the scanner. Do not forcefully pull the page out, the scanner will release the page when it has fully exited out of the scanner.

CALIBRATING FROM THE WINDOWS CONTROL PANEL

If you did not install OneTouch during the initial installation of the scanner driver, please follow these steps to open the scanner’s hardware properties.

Note: If you are using Windows Vista 64-bit, or Windows 7 64-bit, please see "Calibrating from TWAIN" on page 103. All other supported versions of Windows can use these instructions or the ones in the TWAIN section.

1.Open the Windows Control Panel.

2.Double-click on Scanners and Cameras.

If you do not see Scanners and Cameras, open Printers and Other Hardware then you should see this option.

3.The Strobe 400 should be listed as one of the attached scanners. Right-click on the scanner and choose Properties to open the Device Settings for your scanner.

4.Insert the Calibration page in the scanner, in the direction of the arrow(s) on the page.

The calibration page can be inserted in either direction, as long as one of the arrows is pointing in towards the scanner.

5.When calibration is complete the scanner will eject the page out of the scanner. Do not forcefully pull the page out, the scanner will release the page when it has fully exited out of the scanner.

Page 104
Image 104
Visioneer 400 manual Calibrating from the Windows Control Panel, Double-click on Scanners and Cameras