Counting mode using the ACAI function

The ACAI is a function that improves the accuracy of the unit mass automatically by increasing the number of samples as the counting process.

ACAI: Automatic Counting Accuracy Improvement

8If a few more samples are added, the processing indicator turns on. To prevent an error, add three or more. The processing indicator does not turn on if overloaded. Try to add the same number of samples as displayed.

9The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch the balance or samples on the pan until the processing indicator turns off.

10Counting accuracy is improved when the processing indicator turns off.

Each time the above operation is performed, a more accurate unit mass will be obtained. There is no definite upper limit to the ACAI range for the number of samples exceeding 100. Try to add the same number of samples as displayed.

11Remove all the samples used in ACAI and proceed with the counting operation using the improved unit mass.

From step 7

Add a few

more samples.

The mark turns on

at proper range.

The mark turns on and off during calculation.

The mark turns off after

the unit mass is improved.

Repeat

Note ACAI will not function on the unit mass entered using the keys, or digital input mode.