7.Response Adjustment

This function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and/or vibration at the place where the balance is installed. This function adjusts by automatically analyzing the environment or by hand-operation. The function has three stages as follows :

Indicator

Parameter

Response

Stability

FAST

Cond 0

Fast response,

Sensitive value

MID.

Cond

1

Slow response, Stable value

SLOW

Cond

2

Response indicator

7.1.Automatic Response Adjustment

This function automatically updates the response adjustment by analyzing the influence of the environment using the internal mass.

1 Press and hold the MODE key until RESPONSE is displayed, then release the key.

2The balance automatically sets the response characteristic.

Caution Do not allow vibration or drafts to affect the balance during adjustment.

3After automatic adjustment, the balance displays end , returns to the weighing mode and displays the updated response indicator for about thirty seconds.

Press and hold

Release

Result

Note

If the automatic response adjustment fails, the balance displays CH ng . Check the ambient

conditions such as breeze and vibration, also check the weighing pan. Then, perform the

adjustment again. To return to the weighing mode, press the

 

 

key.

CAL

If there is matter on the weighing pan, the balance displays

 

 

 

. Remove them from the

 

CH 0

 

pan. To return to the weighing mode, press the

 

key.

 

 

 

CAL

 

 

 

Advise

If the automatic response adjustment is not helpful, try "7.2. Manual Response Adjustment".

21

Page 23
Image 23
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Automatic Response Adjustment, Advise