12.2.Data Memory for Weighing Data

The balance can store 200 sets of weighing data in memory (if time and date are added, the balance can store 100 sets). Even if the AC adapter is removed, the data is maintained in non-volatile memory.

It is not necessary to connect the printer or personal computer to the balance continually, because the balance stores the weighing data in memory.

The data in memory can be output at one time to a printer or personal computer. The data in memory can be displayed on the balance for confirmation.

Data (ID number, data number, time and date) to be added to the output data can be selected in the function setting.

Storing the weighing data

Note If "Clr" blinks in the upper left of the display, delete the data in memory.

1Set the "Data memory (data)" parameter to "1".

2Specify the "Time/Date output (5-td)" parameter whether time and date is to added or not.

3The storing mode depends on the "Data output mode (prt)" parameter setting. Four types of storing modes are available to store data.

Key mode........................ When the PRINT key is pressed and the displayed value is

stable, the balance stores the weighing data.

Auto print modes A ......... When the displayed value is stable and the conditions of "Auto

print polarity", "Auto print difference" and zero point (reference

value) are met, the balance stores the weighing data.

Auto print modes B ......... When the displayed value is stable and the conditions of "Auto

print polarity", "Auto print difference" and last stable data (reference value) are met, the balance stores the weighing data.

Interval memory mode.... Weighing data is stored at an interval specified in "Interval time

(int)". Press the

PRINT key to start and stop this mode.

The data amount

 

Interval memory

in memory

Weighing data

Memory full

standby indicator

being displayed

 

 

Interval memory

Data number of the data

 

 

active indicator

currently displayed

 

 

 

Memory data

 

 

being displayed

 

Caution

When weighing data is being stored in memory, the data can not be output to a personal computer using the RS-232C interface.

"fUl" indicates that memory is full or the memory capacity has been reached. More data can not be stored unless the memory data is deleted.

53

Page 55
Image 55
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Data Memory for Weighing Data, Storing the weighing data, Memory full