Entering the density of a liquid

Two ways to set the density of a liquid are available in the function table, "Liquid density input (ldin): by entering the water temperature or by entering the density directly.

1 Press the

 

key as necessary to select the density mode. When

MODE

the density mode begins, "g (gram)" is displayed and the processing

indicator (

) blinks.

Entering the water temperature (ldin 0)

2 The water temperature currently set (unit:°C, factory setting : 25°C) is displayed. Use the following keys to change the value.

RE-ZERO(+)key .. The key to increase the temperature by one degree. (0°C is displayed after 99°C)

MODE

........(-)key

The key to decrease the temperature by one degree.

 

 

(99°C is displayed after 0°C)

PRINT

key

The key to store new water temperature, display

 

return to the density mode. Proceed to Step 1.

end

and

CAL

key

The key to cancel the change and return to the density mode.

 

Proceed to Step 1.

The relation between the water temperature and density

°C

+0

+1

+2

+3

+4

+5

+6

+7

+8

+9

0

0.99984

0.99990

0.99994

0.99996

0.99997

0.99996

0.99994

0.99990

0.99985

0.99978

10

0.99970

0.99961

0.99949

0.99938

0.99924

0.99910

0.99894

0.99877

0.99860

0.99841

20

0.99820

0.99799

0.99777

0.99754

0.99730

0.99704

0.99678

0.99651

0.99623

0.99594

30

0.99565

0.99534

0.99503

0.99470

0.99437

0.99403

0.99368

0.99333

0.99297

0.99259

40

0.99222

0.99183

0.99144

0.99104

0.99063

0.99021

0.98979

0.98936

0.98893

0.98849

50

0.98804

0.98758

0.98712

0.98665

0.98618

0.98570

0.98521

0.98471

0.98422

0.98371

60

0.98320

0.98268

0.98216

0.98163

0.98110

0.98055

0.98001

0.97946

0.97890

0.97834

70

0.97777

0.97720

0.97662

0.97603

0.97544

0.97485

0.97425

0.97364

0.97303

0.97242

80

0.97180

0.97117

0.97054

0.96991

0.96927

0.96862

0.96797

0.96731

0.96665

0.96600

90

0.96532

0.96465

0.96397

0.96328

0.96259

0.96190

0.96120

0.96050

0.95979

0.95906

Entering the density directly (ldin 1)

The density currently set (unit : g / cm3, factory setting : 1.0000g / cm3) is displayed. Use the following keys to change the value.

The range to set the density is 0.0000g / cm3 to 1.9999g / cm3. RE-ZERO(+)key.. The key to set the value of the digit selected.

RANGE (-)key......The key to select the digit to change the value.

PRINT

key ......... The key to store the change, display

mode. Proceed to Step 1.

end

and return to the density

CAL

key............. The key to cancel the change and return to the density mode. Proceed to

Step 1.

59