17. Specifications

 

 

 

 

GH-120

GH-200

GH-300

 

GH-202

 

GH-252

Weighing capacity

120 g

220 g

320 g

 

220 g

 

250 g

 

51 g

 

101 g

 

 

 

 

 

 

 

 

 

Maximum display

120.0084 g

220.0084 g

320.0084 g

 

220.0084 g

 

250.0084 g

 

51.00009 g

 

101.00009 g

 

 

 

 

 

 

 

 

 

Minimum display

 

0.1 mg

 

 

0.1

mg

 

 

 

0.01 mg

 

 

 

 

 

 

 

 

Repeatability

0.1 mg

0.2mg

 

0.1 mg

(Standard deviation)

 

0.02 mg

 

0.03 mg

 

 

 

 

 

Linearity

±0.2 mg

±0.3 mg

 

±0.2

mg

 

±0.03 mg

 

±0.10 mg

 

 

 

 

 

 

 

 

 

Stabilization time

Approx. 3.5 seconds

 

Approx. 3.5 seconds

 

(Typical at

 

)

 

Approx. 8 seconds

 

FAST

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensitivity drift,

 

 

±2 ppm/°C

 

 

 

10°C to 30°C / 50°F to 86°F

 

 

 

 

 

 

 

 

 

 

 

 

Operating environment

5°C to 40°C, 85%RH or less (No condensation)

Display refresh rate

 

5 times/second or 10 times/second

 

 

Counting mode

Minimum unit

 

 

0.1 mg

 

 

 

mass

 

 

 

 

 

 

 

 

 

 

 

 

Number of

 

10, 25, 50 or 100 pieces

 

 

samples

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Percent mode

Minimum 100%

 

 

10.0 mg

 

 

 

reference mass

 

 

 

 

 

 

 

 

 

 

 

 

Minimum 100%

0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)

 

display

 

 

 

 

 

 

 

Interface

 

RS-232C with Windows Communication Tools Software WinCT

 

 

 

 

 

 

300g

 

 

 

 

External calibration

 

200g

200g

 

200g

100g

100g

100g

 

100g

weight

 

 

 

 

 

 

 

50g

 

 

 

 

50g

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20g

Weighing pan

 

 

φ90 mm

 

 

 

 

External dimensions

 

217(W) x 442(D) x 316(H) mm

 

 

Power supply &

Power consumption: Approx. 11VA (supplied to the AC adapter )

Confirm that the adapter type is correct for the local voltage and

AC adapter type

 

power receptacle type.

 

 

 

 

 

 

 

 

 

Mass

 

 

 

 

 

Approx. 8.2 kg

 

 

 

The operating environment does not include excessive change of ambient temperature, humidity, vibration, drafts, magnetic fields and static electricity.

The internal mass may change due to corrosion or other damage caused by the operating environment, or due to aging. Check the internal mass periodically and correct the internal mass value if necessary.

74

Page 76
Image 76