9.2.Initializing the Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
External calibration weight and target mass value
Function switch settings ("9.1. Permit Or Inhibit") Liquid density and temperature in the density mode
Note Be sure to calibrate the balance after initialization.
1 Press the
ON:OFF
key to turn off the display.
2 While pressing and holding the PRINT key and the RANGE key, press the ON:OFF key to display p5
. | Press and hold Press |
3 Press the
RANGE
key to display
Clr
.
4Press the PRINT key.
To cancel this operation, press the
CAL
key.
5 Press the
key to display
Clr go
.
6 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.