9.Function Switch and Initialization

9.1.Permit or Inhibit

The balance stores parameters that must not be changed unintentionally (Example: Calibration data for accurate weighing, Data for adapting to the operating environment, Control data for the RS-232C interface). There are five switches for the purpose of protecting parameters. Each switch can select either "permit" or "inhibit". "Inhibit" protects parameters against unintentional operations.

1 Press the

ON:OFF

key to turn off the display.

2 While pressing and holding the PRINT press the ON:OFF key to display p5

key and the

.

RANGE

key,

3 Press the

PRINT

key. Then the balance displays the function switches.

4Specify the switches using the following keys. RANGE key ...........The key to select blinking digit.

RE-ZEROkey.........The key to change the parameter for the selected switch. 0 To inhibit changes. (Can not be used.)

1 To permit changes. (Can be used.)

PRINT key .............The key to store the new parameter and return to the weighing mode.

.................

The key to cancel current operation and return to the weighing mode.

CAL key

Function table

0To inhibit changes to the function table.

1To permit changes to the function table.

Calibration using the internal mass (One-touch calibration)

0To inhibit calibration using the internal mass.

1To permit calibration using the internal mass.

Calibration using the external weight

0To inhibit calibration using the external weight.

1To permit calibration using the external weight.

Automatic self calibration (due to changes of ambient temperature)

0To inhibit automatic self calibration.

1To permit automatic self calibration.

Internal mass correction

0To inhibit correction.

1To permit correction.

29