AD-1653 Density Determination Kit

Weigh in air

x water density = sample density

Weigh in water - Weigh in the air

Example

 

 

 

 

 

 

10.0000 g

 

 

 

 

x 0.9970 g/cm3

= 21.4 g/cm3

 

 

10.0000 g 9.5334 g

 

 

 

 

 

 

 

 

Temperature

Water density

 

 

 

0°C

0.99984 g/cm3

 

 

 

10°C

0.99970 g/cm3

 

 

 

20°C

0.99820 g/cm3

 

 

 

30°C

0.99565 g/cm3

 

 

AD-1683 DC Static Eliminator

The eliminator is able to dissipate static electricity on a charged sample.

The eliminator does not use a ventilator and can be used in precision powder weighting.

AX-USB-9P-EX USB Converter

Adds a COM port to a PC.

Enables bi-directional communication between the PC and the balance when a USB driver is installed.

Can use serial communication software such as WinCT on a PC without COM ports. An RS-232C cable is provided to connect the USB converter to the balance.

RS-232C

on the GH series

balance

USB serial

converter

To USB port

on a PC

RS-232C cable provided (D-sub 9-pin)

77

Page 79
Image 79
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual AD-1653 Density Determination Kit, AD-1683 DC Static Eliminator, 9970 g/cm3