8.Calibration
8.1.Calibration Group
Calibration
Automatic self calibration (Calibration due to changes in temperature)
Calibration using the internal mass
Calibration using an external weight that you have
Calibration test
Calibration test using the internal mass
Calibration test using target mass that you have
Correction of the internal mass value
Correction of the internal mass value
Caution
Do not allow vibration or drafts to affect the balance during calibration. Calibration test does not perform calibration.
To output the data for GLP using the
Calibration test is available only when "GLP output (info)" of "Data output (dout)" is set . The calibration and calibration test data can be stored in memory. To store them, set "Data memory (data)". Refer to "12. Data Memory" for details.
Caution on using an external weight
The accuracy of an external weight can influence the accuracy of weighing. Select a mass for calibration and calibration test from the following table.
Model |
| Usable calibration mass |
| Adjustable range | ||
| 50g, | 100g* |
|
| ||
|
| 100g, | 200g* |
| ||
|
| 100g, | 200g*, | 300g |
| |
20g, | 50g, | 100g, | 200g* |
|
The calibration mass in bold type: factory setting
The calibration mass value can be adjusted within the range above.
Display
This indicator means "In process of measuring calibration data".
Do not allow vibration or drafts to affect the balance while the indicator is displayed.
About the internal mass
The internal mass may change due to corrosion or other damage caused by the operating environment, or due to aging. Check the internal mass periodically and correct the internal mass value if necessary.