12.3.Data Memory for Calibration and Calibration Test

Calibration data (when and how it is performed) and calibration test

 

data can be stored in memory.

 

All the data in memory is available to be output at one time to a

 

printer or personal computer.

Indicator

Up to 50 data sets of the latest calibration or calibration test can be

stored. When the memory capacity has been reached, "fUl"

 

illuminates in the upper left of the display as shown below.

 

Storing the calibration and calibration test data

Note If "Clr" appears blinking in the upper left of the display, delete the data in memory.

Store the calibration and calibration test data as follows:

1Set the "Data memory (data)" parameter to "2".

2Set the "GLP output (info)" parameter to "1" or "2".

3With the settings above, each time calibration or calibration test is performed, the data is stored automatically.

Transmitting the memory data

Note

Confirm that the "Serial interface (5if)" parameters are set properly.

Refer to "10. Function Table" and "15.2. Connection To Peripheral Equipment". Confirm that the "Data memory (data)" parameter is set to "2".

1

Press and hold the

 

 

 

key until

out

 

is displayed, then

 

PRINT

 

release the key.

 

 

 

 

 

 

 

 

 

 

2

Press the

 

key to display

 

 

 

 

with "no" blinking.

PRINT

out "n"o

 

 

3

Press the

 

 

key to display

 

 

 

with "go" blinking.

 

 

 

 

 

out

"g"o

RE-ZERO

 

In the upper left

of the display

4 Press the

PRINT

key to transmit all memory data using the RS-232C interface.

5 The balance displays Clear when all memory data is output. Press the CAL key to return to the weighing mode.

Deleting data stored in memory

1 Press and hold the

PRINT

key until

out

is displayed, then release the key.

2Press the

3Press the

4Press the

5Press the

 

 

 

key to display

 

Clear

.

 

In the upper left

SAMPLE

 

 

key to display

 

 

 

 

 

with "no" blinking.

of the display

PRINT

Clr "n"o

 

 

 

 

 

 

 

key to display

 

 

 

with "go" blinking.

 

 

 

 

 

Clr

"g"o

 

RE-ZERO

 

 

PRINT key to delete all data.

6The balance displays out and returns to the weighing mode when all the data has been deleted.

56

Page 58
Image 58
A&D GH-120/GH-200/GH-300/GH-202/GH-252 Data Memory for Calibration and Calibration Test, Transmitting the memory data