Upper left of the display

12. Data Memory

Data memory is a function to store weighing data and calibration data in memory. The data stored in memory are available for outputting at one time to a printer or personal computer.

Weighing data

Excluding date and time

Up to 200 sets

Including date and time

Up to 100 sets

 

Calibration report

 

 

Internal calibration

 

External calibration

Last 50 sets

Calibration test report

 

Internal test calibration

 

External test calibration

 

12.1.Notes on Using Data Memory

To use the memory function, set the "Data memory (data)" parameter and "Time/Date output (5-td)" parameter of the function table. Refer to "10. The Function Table" for details on setting the data memory.

For weighing data, the data contents to be stored and the storage capacity depend on the "Time/Date output (5-td)" parameter setting.

If a different type of data exists in memory, "Clr" blinks the upper left of the display. For example, you want to store weighing data but calibration data or unit mass data remains in memory.

Under such a condition, before storing data, delete the data in memory as follows:

Releasing "Clr" or "err"

1Press and hold the PRINT key until the key.

SClr "n"o

with "no" blinking is displayed, then release

2Press the RE-ZEROkey to display SClr "g"o with "go" blinking.

The type of data stored in memory appears in the upper left of the display as shown below:

Weighing data without time and date

-d-

Weighing data with time and date

d-t

Calibration report

Hi5

3 Press the

PRINT

key to delete all the data in memory.

4 The balance displays

end

and returns to the weighing mode.

52