15.3.Commands
15.3.1.Command List
Note A command has a terminator added, that is specified using "5if Crlf" of the function table, and is sent to the balance.
| Commands to query weighing data |
|
|
|
|
|
| |||||||
| C |
|
|
|
| Cancels the S or SIR command. |
| |||||||
| Q |
|
|
|
| Requests the weighing data immediately. |
| |||||||
| S |
|
|
|
| Requests the weighing data when stabilized. |
| |||||||
| SI |
|
|
|
| Requests the weighing data immediately. |
| |||||||
| SIR |
|
|
|
| Requests the weighing data continuously. |
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| Commands to control the balance |
|
|
|
|
|
| |||||||
| CAL |
|
|
|
| Same as the |
|
| CAL | key. |
| |||
|
|
|
|
|
|
|
|
|
|
|
| |||
| OFF |
|
|
|
| Turns the display off. |
|
|
| |||||
| ON |
|
|
|
| Turns the display on. |
|
|
| |||||
| P |
|
|
|
| Same as the |
|
|
|
| key |
| ||
|
|
|
|
|
| ON:OFF |
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| PRT |
|
|
|
| Same as the |
|
|
|
| key |
| ||
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| R |
|
|
|
| Same as the |
|
|
|
|
| key (Note 1) |
| |
|
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
| |||
| RNG |
|
|
|
| Same as the |
|
|
|
| key. |
| ||
|
|
|
|
|
| RANGE |
| |||||||
|
|
|
|
|
|
|
|
| ||||||
| TR |
|
|
|
| Tares the balance (Cancels the container’s weight.) (Note 1) |
| |||||||
| TST |
|
|
|
| Perform calibration test. |
| |||||||
| U |
|
|
|
| Same as the |
|
|
|
| key |
| ||
|
|
|
|
|
| MODE |
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| ||||
| Commands to request data |
|
|
|
|
|
| |||||||
| ?ID |
|
|
|
| Requests the identification number. |
| |||||||
| ?PT |
|
|
|
| Outputs the tare value. (Note 1) |
| |||||||
| ?SN |
|
|
|
| Request the serial number of the balance. |
| |||||||
| ?TN |
|
|
|
| Request the model name of the balance. |
| |||||||
|
|
|
|
|
|
|
|
|
|
| ||||
| Command to set data |
|
|
|
|
|
| |||||||
| ∗∗∗ ∗∗∗∗ |
|
|
| g | Sets the tare value. (Note 1) |
| |||||||
| PT: . |
|
|
|
|
|
|
|
|
|
|
|
|
(Note 1) | R command assumes the point as zero and sets the display to zero. |
| |
| TR command cancels the tare weight when the displayed value is greater than zero. |
| The canceled tare value can be read using the ?PT command. |
| PT: command sets the tare value digitally. |
| An example of this command sets a negative target value and fills with the sample |
| until the display becomes zero. |
| Step 1 Place a container. |
| Step 2 Set the display to zero using the R command. |
| Step 3 Set a target value using the PT: command. |
| For details, refer to page 69. |
64