16.Maintenance
16.1.Treatment of the Balance
Clean the balance with a lint free cloth that is moistened with warm water and a mild detergent.
Do not use organic solvents to clean the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Use the original packing material for transportation. Consider "3. Precautions" when operation the balance.
16.2.Error Codes
Display | Error code | Description |
|
|
|
| Stability error |
|
|
|
| The balance can not stabilize due to an environmental | ||
| EC,E11 | problem. Check around the pan. Prevent vibration, drafts, | ||
|
| temperature changes, static electricity and magnetic fields, | ||
|
| from influencing the balance. |
| |
|
| To return to the weighing mode, press the CAL key. | ||
|
| Internal mass error |
|
|
| EC,E16 | Applying the internal mass does not yield a change in the | ||
| mass value as specified. |
|
| |
|
| Confirm that there is nothing on the pan and perform the | ||
|
| weighing operation from the beginning again. | ||
|
| Internal mass error |
|
|
| EC,E17 | The internal mass application mechanism does not function | ||
|
| properly. Perform the weighing operation from the beginning | ||
|
| again. |
|
|
|
| Calibration weight error |
|
|
| EC,E20 | The calibration weight is too heavy. Confirm the calibration | ||
|
| mass value. Press the | CAL | key to return to the weighing |
|
| mode. |
|
|
|
| Calibration weight error |
|
|
| EC,E21 | The calibration weight is too light. Confirm the calibration | ||
| mass value. Press the | CAL | key to return to the weighing | |
|
| mode. |
|
|
Overload error
A sample beyond the balance weighing capacity has been placed on the pan. Remove the sample from the pan.
Weighing pan Error
The mass value is too light.
Confirm that the weighing pan is properly installed and calibrate the balance.