16.Maintenance

16.1.Treatment of the Balance

Clean the balance with a lint free cloth that is moistened with warm water and a mild detergent.

Do not use organic solvents to clean the balance.

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.

Use the original packing material for transportation. Consider "3. Precautions" when operation the balance.

16.2.Error Codes

Display

Error code

Description

 

 

 

 

Stability error

 

 

 

 

The balance can not stabilize due to an environmental

 

EC,E11

problem. Check around the pan. Prevent vibration, drafts,

 

 

temperature changes, static electricity and magnetic fields,

 

 

from influencing the balance.

 

 

 

To return to the weighing mode, press the CAL key.

 

 

Internal mass error

 

 

 

EC,E16

Applying the internal mass does not yield a change in the

 

mass value as specified.

 

 

 

 

Confirm that there is nothing on the pan and perform the

 

 

weighing operation from the beginning again.

 

 

Internal mass error

 

 

 

EC,E17

The internal mass application mechanism does not function

 

 

properly. Perform the weighing operation from the beginning

 

 

again.

 

 

 

 

Calibration weight error

 

 

 

EC,E20

The calibration weight is too heavy. Confirm the calibration

 

 

mass value. Press the

CAL

key to return to the weighing

 

 

mode.

 

 

 

 

Calibration weight error

 

 

 

EC,E21

The calibration weight is too light. Confirm the calibration

 

mass value. Press the

CAL

key to return to the weighing

 

 

mode.

 

 

Overload error

A sample beyond the balance weighing capacity has been placed on the pan. Remove the sample from the pan.

Weighing pan Error

The mass value is too light.

Confirm that the weighing pan is properly installed and calibrate the balance.