Calibration test report using an external weight

Note Calibration test does not perform calibration.

Key operation

 

 

 

 

1

 

Press and hold the

 

key to display

 

CC out

and release the key.

CAL

2

 

 

 

is displayed.

 

 

 

 

 

Cal 0

 

 

 

 

3

When updating the target value, press the

 

 

key and proceed to step 4.

 

RANGE

 

 

When using preset target value, proceed to step 5.

4

Specify calibration mass value using the following keys.

 

 

 

 

key ........ The key to select the blinking figure

 

 

RANGE

 

 

 

 

 

(+)key .. The key to increase the value of the blinking figure.

 

 

RE-ZERO

 

 

 

 

(-)key........The key to decrease the value of the blinking figure.

 

 

MODE

 

 

 

key ......... The key to store the new value.

 

 

PRINT

5When pressing the RE-ZEROkey, the zero point is measured and the weight value is displayed for a few seconds.

6Place the displayed mass on the pan. Press the PRINT key to measure the mass and the weight value is displayed for a few seconds.

7

If GLP output is used,

glp

is displayed and the calibration test report is output.

8

The balance returns to weighing mode after this test.

Setting of "info 1" AD-8121 printer format

-----------A-&-D

MODEL-----GH-300

S/N-----01234567

ID------LAB-0123

DATE--2004/07/01 TIME----12:34:56

CAL.TEST(EXT.)

ACTUAL

-------0.0000--g

----+200.0002--g

TARGET

----+200.0000--g

SIGNATURE

Setting of "info 2" General format

Manufacturer

Model

Serial number

ID number

Date

Time Calibration test type

Zero point value

Target weight value

Target weight

Signature

---------------

 

 

 

 

Space, ASCII 20h

 

 

 

 

 

 

 

 

 

 

 

 

 

<TERM>

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

50

Page 52
Image 52