Interval memory mode

The weighing data is periodically stored in memory.

Example

For periodical weighing without a personal computer command and

 

to output all of the data, to a computer, at one time.

 

The GH series can use time and date with "Time/Date output (5-td)".

Required setting

dout

prt 3

Interval memory mode

 

dout

data 2

Data memory function is used

 

dout

int

Interval time

Optional setting

dout

5-td1,2, or 3

Adds the time and date.

10.5.Description of the Item "Data Format"

A&D standard format

5if type 0

This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is used, set the printer to MODE 1 or 2.

This format consists of fifteen or sixteen characters excluding the terminator.

When numerical characters without decimal point are exceeded eight characters for the GH-252, the format becomes sixteen characters.

A header of two characters indicates the balance condition.

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is applied.

The unit, consisting of three characters, follows the data.

S

T , +

0 0 0 .

1 2 7 8

g

CR LF

Header Data

ST Stable header U S Unstable header O L Overload header

Q T

Unit Terminator

Stable header of counting mode

S T , + 1 0 0

.

0

1

2

7

8

gCR LF

Header

Data

Unit

Terminator

DP (Dump print) format

5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B is used, set the printer to MODE 3.

This format consists of sixteen characters excluding the terminator.

A header of two characters indicates the balance condition. No overload header is used. The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero or overloaded.

The unit, consisting of three characters, follows the data.

W

T

 

 

 

 

+

0

.

1

2

 

7

 

8

 

Header

 

 

Data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Stable header

 

 

 

 

 

 

W

T

 

 

Q

 

T

 

US Unstable header

gCR LF

Unit Terminator

Stable header of counting mode

39

Page 41
Image 41
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Interval memory mode, Description of the Item Data Format, Standard format