Display

Error code

Description

 

 

Sample mass error

 

 

The balance can not store the sample for the counting mode

 

 

or for the percent mode because it is too light.

 

 

Use a larger sample.

 

 

Unit mass error

 

 

The sample unit mass for the counting mode is too light.

 

 

Storing and using it for counting will cause a counting error.

 

 

Add samples to reach the specified number and press the

 

 

PRINT

key. Pressing the PRINT key without adding

 

 

samples will shift the balance to the counting mode. But, for

 

 

accurate counting, be sure to add samples.

 

 

Automatic response adjustment zero error

 

 

The automatic response adjustment can not be performed

 

 

because there is something on the pan. Clear the pan. Press

 

 

the CAL

key to return to the weighing mode.

 

 

Automatic response adjustment unstable error

 

 

The automatic response adjustment can not be performed

 

 

because the mass value is unstable.

(Check NG)

Check the ambient conditions such as breeze, vibration and

magnetic fields, also check the weighing pan.

 

 

 

 

Press the

CAL key to return to the weighing mode.

 

 

Clock battery error

 

 

The clock backup battery has been depleted.

 

 

Press any key and set the time and date. The clock and

 

 

calendar function works normally as long as the AC adapter

 

 

is connected to the balance. If this error appears frequently,

 

 

contact the local A&D dealer.

 

 

Memory full

"f"U"l

(Blinking)

The amount of weighing data in memory has reached the

maximum capacity. Delete data in memory to store new data.

 

 

 

For details, refer to "12. Data Memory".

 

 

Memory full

 

 

The amount of calibration or calibration test data in memory

fU l

(Illuminated)

has reached the maximum capacity (50 sets).

 

The data in memory will be deleted automatically to store new

 

 

 

 

data. For details, refer to "12. Data Memory".

 

 

Memory type error

"C"l"r

 

Type of memory set in the function table and type of data

 

 

stored are different. For details, refer to "12. Data Memory".

 

 

Memory data error

 

 

The stored data were destroyed. To use the memory

 

 

function, erase all the old data and release the error. For

 

 

details, refer to "12.1. Notes on Using Data Memory".