Automatic self calibration can not be used while the interval memory mode is active. The following commands can not be used during data storage.

QQuery command for weighing data.

SQuery command for stable weighing data. SI Query command for weighing data.

SIR Query command for continuous weighing data.

Setting the function table

Parameter settings for each output mode are as follows:

Item

Data output

Auto print polarity,

Data memory

Interval time

Mode

mode

difference

function

 

Key mode

prt 0

Not used

data 1

Not used

Auto print mode A

prt 1

ap-a 0-2

data 1

Auto print mode B

prt 2

ap-b 0-2

data 1

 

Interval memory mode

prt 3

Not used

data 1

int 0-8

Additional parameter settings, as follows:

Data number

No

d-no 0

Yes

d-no 1

 

ID number

No

5-id 0

Yes

5-id 1

 

 

No

5-td 0

Time and date

Time only

5-td 1

Date only

5-td

2

 

 

Both

5-td

3

Enabling the data memory function

1

Press and hold the

 

 

 

key until

ba5fnc

 

is displayed, then release the key.

 

RANGE

2

Press the

 

 

 

 

key several times to display

 

 

.

 

 

 

 

 

dout

RANGE

3

Press the

 

 

key.

 

 

 

 

PRINT

 

 

 

 

4

Press the

 

 

 

key three times to display

 

 

.

 

 

 

 

data 0

 

RANGE

5

Press the

 

 

 

key to display

 

.

 

 

 

 

 

 

 

 

 

 

data 1

 

 

 

 

RE-ZERO

 

 

 

 

 

6

Press the

 

 

key to store the setting.

 

 

 

 

PRINT

 

 

 

 

7

Press the

 

key to return to the weighing mode.

CAL

54

Page 56
Image 56
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Setting the function table, Enabling the data memory function