10.2.Details of the Function Table

Class

Item and Parameter

 

 

Description

 

Cond

 

0

Fast response, sensitive value

 

FAST

 

Can be changed by response

 

1

 

 

 

 

Condition

 

 

 

 

MID.

 

adjustment.

 

2

Slow response, stable value

 

 

 

 

SLOW

 

 

 

 

 

 

 

 

5t-b

 

0

Stable range is ±1 digit

 

 

 

 

The stabilization indicator illuminates

 

 

1

 

#1

 

when the display fluctuation is within

 

Stability band width

 

 

 

 

 

2

Stable range is ±3 digits

 

 

 

 

the range.

 

 

 

 

 

 

 

 

 

 

ba5fnc

 

 

0

OFF

 

 

 

 

 

 

 

trc

 

1

Normal

 

 

 

 

Keeps zero display by tracking zero

Environment

 

 

 

 

 

Zero tracking

 

2

Strong

 

 

 

 

drift.

 

 

 

 

 

Display

 

 

3

Very strong

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5pd

 

0

5 times/second

 

 

 

 

Period to refresh the display

 

Display refresh rate

 

1

10 times/second

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pnt

 

0

Point (.)

 

 

 

 

Decimal point format

 

Decimal point

 

1

Comma (,)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

p-on

 

0

OFF

 

 

 

 

Turns on the weighing mode display

 

Auto display-ON

 

1

ON

 

 

 

 

when the AC adapter is connected.

Cl adj

Refer to "10.7. Clock and Calendar Function"

 

 

 

 

The time and date are added to the

Clock

 

 

 

 

output data.

 

 

 

 

 

 

 

 

 

 

 

0

Key mode

 

 

 

 

Accepts the

PRINT

key only when

 

 

 

 

 

 

 

the display is stable.

 

 

 

 

 

 

 

 

 

 

 

 

1

Auto print mode A

 

 

 

 

Outputs data when the display is

 

prt

 

 

(Reference = zero)

 

 

 

 

 

 

 

 

 

 

 

stable and conditions of ap-p, ap-b

 

Data output mode

 

 

Auto print mode B

 

 

 

 

 

 

2

 

 

 

 

and the reference value are met.

 

 

 

(Reference = last stable value)

 

With data 0, outputs data

 

 

 

 

Stream mode /

 

 

 

 

 

 

 

3

 

 

 

 

continuously; with data 2, uses

dout

 

 

 

Interval memory mode

 

 

 

 

interval memory.

 

 

0

Plus only

 

 

 

 

Displayed value>Reference

Data output

ap-p

 

 

 

 

 

 

1

Minus only

 

 

 

 

Displayed value<Reference

 

Auto print polarity

 

 

 

 

 

 

 

2

Both

 

 

 

 

Regardless of displayed value

 

 

 

 

 

 

 

 

ap-b

 

0

10 digits

 

 

 

 

Difference between reference value

 

 

1

100 digits

 

 

 

 

and displayed value

 

Auto print difference

 

 

 

 

 

 

 

2

1000 digits

 

 

 

#2

 

 

 

 

 

 

 

 

 

 

 

 

 

data

 

0

Not used

 

 

 

 

Related items:

 

 

1

Stores weighing data

 

 

 

 

 

Data memory

 

 

 

 

 

prt,int, d-no, 5-td, info

 

 

2

Stores calibration data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

: Factory settings.

#1 The unit of minimum display is digit.

Example: If 1 mg display is selected using the RANGE key for the GH-300, 1 mg is one digit.

#2 Usable minimum display of the balance is one digit.

Example: In gram display, one digit is 0.01 mg for the GH-252 and 0.1 mg for the GH-300.

34

Page 36
Image 36
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Details of the Function Table