Title block and end block

When mass values are recorded as GLP data, a "Title block" is inserted at the beginning and an "End block" is inserted at the end of a group of mass values in the GLP report.

Notes

To output the report to an AD-8121B , use MODE 3 of the AD-8121B. If MODE1 is used, select temporary dump print mode by pressing the STAT. key of the AD-8121B.

If the data memory function is used (except data 0), the "Title block" and "End block" can not be output.

Key operation

1 With the weighing data displayed, press and hold the PRINT key, then release the key. 5tart is displayed. The "Title block" is output.

2The weighing data is output according to the parameter setting of the data output mode (prt) of the function table.

3 Press and hold the PRINT key until recend is displayed, then release the key. The "End block" is output.

Setting of "info 1" AD-8121 printer format

A & D

MODEL GH-300

S/N 01234567

ID LAB-0123 DATE 2004/07/01

START

TIME 12:34:56

WT +123.4567 g WT +123.4612 g WT +123.4623 g WT +123.4634 g WT +123.4511 g WT +123.4534 g WT +123.4712 g WT +123.4645 g

Title block

Setting of "info 2"

General format

Manufacturer

Model

Serial number

ID number

Date

Time

Weighing data

END

TIME 12:40:15

SIGNATURE

- - - - - - - -

Time

Signature

 

 

 

 

End block

 

 

 

 

Space, ASCII 20h

 

 

 

 

<TERM>

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

 

 

 

 

51

Page 53
Image 53
A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Title block and end block, Weighing data