Calibration
8-
6
ADAM 4000 Series User’s Manual
Table 8-2 Calibration Resistance
Module Input
Range
Code
(Hex)
Input Range Span
Calibration
Resistance
Offset
Calibration
Resistance
4013 20 Pt, -100~100 ℃
a = 0.00385
140
Ω
60 Ω
21 Pt, 0~100 ℃
a = 0.00385
140
Ω
60 Ω
22 Pt, 0~200 ℃
a = 0.00385
200
Ω
60 Ω
23 Pt, 0~600 ℃
a = 0.00385
440
Ω
60 Ω
24 Pt, -100~100 ℃
a = 0.003916
140
Ω
60 Ω
25 Pt, 0~100 ℃
a = 0.003916
140
Ω
60 Ω
26 Pt, 0~200 ℃
a = 0.003916
200
Ω
60 Ω
27 Pt, 0~200 ℃
a = 0.003916
440
Ω
60 Ω
28 Ni, -80~100 ℃ 200
Ω
60 Ω
29 Ni, 0~100 ℃ 200
Ω
60 Ω