Triggering the Oscilloscope

To use TV triggering

Provide Correct Matching

Many TV signals are produced from 75 sources. To provide correct matching to these sources, a 75 terminator (such as an Agilent 11094B) should be connected to the oscilloscope input.

3Press the sync polarity softkey to set the TV trigger to either positive ( ) or negative () sync polarity.

4Press the Standard softkey to set the TV standard.

The oscilloscope supports triggering on NTSC, PAL, PAL-M, SECAM, and Generic television (TV) and video standards. NTSC, PAL, PAL-M, and SECAM are broadcast standards used throughout the world. Generic allows you to trigger on progressive scan and other non-broadcast video signals.

5Press the Mode softkey to select the TV triggering mode

The TV trigger modes available are:

Field1 and Field2 - trigger on the rising edge of the first serration pulse of field 1 or field 2.

All Fields - trigger on the rising edge of the first serration pulse of both field 1 and field 2 (not available in Generic mode).

All Lines - trigger on the first line found.

Line: Field1 and Line:Field2 - trigger on the selected line # in field 1 or field 2.

Line: Alternate - alternately trigger on the selected line # in field 1 and field 2 (not available in Generic mode).

Vertical - trigger on the rising edge of the first serration pulse or ~70 us after the start of vertical sync, whichever occurs first (only available in Generic mode).

Count: Vertical - counts falling edges of sync pulses; triggers on the selected count # (only available in Generic mode).

3-41