Making Measurements

Math Scale and Offset

Math Scale and Offset

Any math function may be manually scaled by pressing the Settings softkey, and then adjusting the Scale or Offset value.

Math Scale and Offset are Set Automatically

Any time the currently displayed math function definition is changed, the function is automatically scaled for optimum vertical scale and offset. If you manually set scale and offset for a function, select a new function, then select the original function, the original function will be automatically rescaled.

1Press the Settings softkey in the Math menu to set your own scale factors (units/division) or offset (units) for the selected math function.

Units for each input channel can be set to Volts or Amps using the channel Probe Units softkey. Scale and offset units are:

Math function

Units

FFT

dB* (decibels)

1*2

V2, A2, or W (Volt-Amp)

1-2

V or A

d/dt

V/s or A/s (V/second or A/second)

dt

Vs or As (V-seconds or A-seconds)

*When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on the 54624A), FFT units

will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 M. FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50(54640-series only). FFT units will be displayed as dB for all other FFT sources or when a source channel’s units has been set to Amps.

A scale unit of U (undefined) will be displayed for math function 1-2 and for d/dt, and dt when 1-2 or 1+2 is the selected source if channel 1 and channel 2 are set to dissimilar units in the channel Probe Units softkey.

2Press the Scale or Offset softkey, then turn the Entry knob to rescale or change the offset value for your math function.

5-18

Page 186
Image 186
Agilent Technologies 54621D, 24A, 42A, 54621A, 22A, 41A manual Math Scale and Offset are Set Automatically